Shape memory alloys in medicine Materiali z oblikovnim spominom v medicini Miha Brojan1, David Bombač 2, Franc Kosel1, Tomaž Videnič 1 'University of Ljubljana, Faculty of Mechanical Engineering, Aškerčeva cesta 6, SI-1000 Ljubljana, Slovenia; E-mail: miha.brojan@fs.uni-lj.si, franc.kosel@fs.uni-lj.si, tomaz.videnic@fs.uni-lj.si 2University of Ljubljana, Faculty of Natural Sciences and Engineering, Aškerčeva cesta 12, SI-1000 Ljubljana, Slovenia; E-mail: david.bombac@ntf.uni-lj.si Received: December 8, 2007 Accepted: April 21, 2008 Abstract: The Shape memmory alloys (SMA) are success story in medical applications market with enormous growth in usage. Huge advances from surgical point of view mean great opportunity for new commercial applications. This paper reviews the development of the shape memory alloys (SMA), constitutive behavior and use in medicine. Shape memory effect, pseudoelasticity and other basic properties of SMA are presented. Later many medical devices using shape memory effect and current commercial applications are presented. Because vast majority of current SMA medical devices is made from Nitinol, this paper also consider the factors that impinge on the associated risk analysis of using Nitinol in medical applications. Izvleček: Materiali z oblikovnim spominom so zgodba o uspehu uporabe v medicini in še vedno pridobivajo na svoji uporabnosti. Velike prednosti s kirurškega stališča pomenijo ogromne možnosti za nove komercialne aplikacije. To delo podaja pregled fizikalnega ozadja materialov z oblikovnim spominom, njihove fizikalne zakonitosti ter uporabnost v medicini. Predstavljeni so pojav oblikovnega spomina, psevdoelastičnost in druge osnovne lastnosti materialov z oblikovnim spominom. V nadaljevanju so predstavljeni medicinski pripomočki, ki izkoriščajo oblikovni spomin in sedanja komercialna uporaba. Ker je večina takšnih medicinskih pripomočkov izdelanih iz materiala nitinol so preučeni tudi vplivi, ki zadevajo nevarnosti uporabe tega materiala v medicinskih aplikacijah. Key words: shape memory effect, shape memory alloys, SMA medical implants, Nitinol Ključne besede: oblikovni spomin, materiali z oblikovnim spominom, SMA medicinski vsadki, nitinol Introduction Smart materials have been given a lot of attention mainly for their innovative use in practical applications. One example of such materials is also the family of shape memory alloys (SMA) which are arguably the first well known and used smart material. Shape memory alloys possess a unique property according to which, after being deformed at one temperature, they can recover to their original shape upon being heated to a higher temperature. The effect was first discussed in the 1930s by Olander[1] and Greninger and Mooradian [2]. The basic phenomenon of the shape memory effect was widely reported a decade later by Russian metallurgist G. Kurdjumov and also by Chang and Read[3]. However, presentation of this property to the wider public came only after the development of the nickel-titanium alloy (nitinol) by Buehler and Wang [4]. Since then, research activity in this field has been intense, and a number of alloys have been investigated, including Ag-Cd, Au-Cd, Cu-Zn, Cu-Zn-Al, Cu-Al-Ni, Cu-Sn, Cu-Au-Zn, Ni-Al, Ti-Ni, Ti-Ni-Cu, Ni-Ti-Nb, Ti-Pd-Ni, In-Ti, In-Cd and others. Crystallography of shape memory alloys have been studied for the last four decades. Only a fraction of the available literature is listed here[5-14]. Because these materials are relatively new, some of the engineering aspects of the material are still not well understood. Many of the typical engineering descriptors, such as young's modulus and yield strength, do not apply to shape memory alloys since they are very strongly temperature dependent. On the other hand, a new set of descriptors must be introduced, such as stress rate and amnesia. That is why numerous constitu- tive models have been proposed over the last 20 years to predict thermomechanical behaviour[15-28]. Thermomechanical behavior These materials have been shown to exhibit extremely large, recoverable strains (on the order of 10 %), and it is these properties as functions of temperature and stress which allow SMAs to be utilized in many exciting and innovative applications. From a macroscopic point of view, the mechanical behavior of SMAs can be separated into two categories: the shape memory effect (SME), where large residual (apparently plastic) strain can be fully recovered upon raising the temperature after loading and unloading cycle; and the pseudoelasticity or superelasticity, where a very large (apparently plastic) strain is fully recovered after loading and unloading at constant temperature. Both effects are results of a martensite phase transformation. In a stress-free state, an SMA material at high temperatures exists in the parent phase (usually a body-centered cubic crystal structure, also refered as the austenite phase). Upon decreasing the material temperature, the crystal structure undergoes a self-accommodating crystal transformation into martensite phase (usually a face-centered cubic structure). The phase change in the unstressed formation of martensite from austenite is referred to as 'self-accommodating' due to the formation of multiple martensitic variants and twins which prohibits the incurrence of a transformation strain. The martensite variants, evenly distributed throughout material, are all crystallographically equivalent, Figure 1. Martensite transformation in shape memory alloys and steels Slika 1. Martenzitna premena v materialih z oblikovnim spominom in jeklih Figure 2. Shape memory effect Slika 2. Pojav spomina oblike differing only by habit plane. The process of self-accomodation by twinning allows an SMA material to exhibit large reversible strains with stress. However, the process of self-accommodation in ordinary materials like stainless steel does not take place by twinning but via a mechanism called slip. Since slip is a permanent or irreversible process, the shape memory effect cannot occur in these materials. The difference between the twinning and slip process is shown in Figure 1. In the stress-free state, an SMA material has four transition temperatures, designated as M, M, A , A, i.e. Martensite Fin- f s> s> f ish, Martensite Start, Austenite Start, and Austenite Finish, respectively. In the case of "Type I" materials, temperatures are arranged in the following manner: M < Ms < A s < A. A change of temperature within the range M s < T < A s induces no phase changes and both phases can coexist within M < T < A. With these four transformation temperatures and the concepts of self-accommodation, the shape memory effect can be adequately explained. As an example let us consider a martensite formed from the parent phase, Figure 2(a), cooled under stress-free conditions through M and M . This material has multiple variants and twins present, Figure 2(b), all crystal-lographically equivalent, but with different orientation (different habit plane indices). When a load applied to this material reaches a certain critical stress, the pairs of martensite twins begin "detwinning" to the stress-prefered twins, Figure 2(c). It means that the multiple martensite variants begin to convert to a single variant determined by alignment of the habit planes with the direction of loading, Figure 2(d). During this process of reorientation, the stress rises very slightly in comparison to the strain. As the single variant of martensite is ther-modynamically stable at T < A , upon unloading there is no conversion to multiple variants and only a small elastic strain is recovered, leaving the material with a large residual strain, Figure 2(e). The detwinned martensite material can recover the entire residual strain by simply heating above A; the material then transforms to the parent phase, which has no variants, and recovers to its original size and shape, Figure 2(f), thus creating the shape memory effect. The pseudoelastic effect can be explained, if an SMA material is considered to be entirely in the parent phase (with T > A.), Figure 3(a). When stress is applied to this material, there is a critical stress at which the crystal phase transformation from austenite to martensite can be induced, Figure 3(b). Due to the presence of stress during the transformation, specific martensite variants will be formed preferentially and at the end of transformation, the stress-induced martensite will consist of a single variant of detwinned martensite, Figure 3(c). During unloading, a reverse transformation to austenite occurs because of the instability of martensite at T > A in the absence of stress, Figure 3(e). This recovery of high strain values upon unloading yields a characteristic hysteresis loop, diagram in Figure 3, which is known as pseudoelastic-ity or superelasticity. Many of the possible medical applications of SMA materials in the 1980's were attempting to use the thermally activated memory effect. However, temperature regions tolerated by the human body are Figure 3. Pseudoelasticity or superelasticity Slika 3. Psevdoelastičnost ali superelastičnost very limited. Small compositional changes around the 50-50 % of Ti-Ni ratio can make dramatic changes in the operating characteristics of the alloy. Therefore very precise control of phase transition temperatures is required. On the other hand, pseu-doelasticity is ideally suited to medical applications since the temperature region of optimum effect can easily be located to encompass ambient temperature through body temperature. Biocompatibility It is important to understand the direct effects of an individual component of the al- loy since it can dissolve in the body due to corrosion and it may cause local and systemic toxicity, carcinogenic effects and immune response. The cytotoxicity of elementary nickel and titanium has been widely researched, especially in the case of nickel, which is a toxic agent and allergen[29-31]. Nickel is known to have toxic effects on soft tissue structures at high concentrations and also appears to be harmful to bone structures, but substantially less than cobalt or vanadium, which are also routinely used in implant alloys. Experiments with toxic metal salts in cell cultures have shown decreasing toxicity in the following order: Co > V > Ni > Cr > Ti > Fe [32]. The dietary exposure to nickel is 160-600 mg/ day. Fortunately most of it is eliminated in the feces, urine and sweat. Pure nickel implanted intramuscularly or inside bone has been found to cause severe local tissue irritation and necrosis and high carcinogenic and toxic potencies. Due to corrosion of medical implants, a small amount of these metal ions is also released into distant organs. Toxic poisoning is later caused by the accumulation, processing and subsequent reaction of the host to the corrosion of the Ni-containing implant. Nickel is also one of the structural components of the metalloproteins and can enter the cell via various mechanisms. Most common Ni2+ ions can enter the cell utilizing the divalent cation receptor or via the support with Mg2+, which are both present in the plasma membrane. Nickel particles in cells can be phagocytosed, which is enhanced by their crystalline nature, negative surface energy, appropriate particle size (2-4 pm) and low solubility. Other nickel compounds formed in the body are most likely to be NiCl2 and NiO, and fortunately there is only a small chance that the most toxic and carcinogenic compounds like Ni3S2, are to be formed. Nickel in soluble form, such as Ni2+ ions, enters through receptors or ion channels and binds to cytoplasmic proteins and does not accumulate in the cell nucleus at concentrations high enough to cause genetic consequences. These soluble Ni2+ ions and are rapidly cleaned from the body. However, the insoluble nickel particles containing phagocytotic vesicles fuse with lysosomes, followed by a decrease of phagocytic in-travesicular pH, which releases Ni2+ ions from nickel containing carrier molecules. The formation of oxygen radicals, DNA damage and thereby inactivation of tumor suppressor genes is contributed by that. On the other hand, titanium is recognized to be one of the most biocompatible materials due to the ability to form a stable titanium oxide layer on its surface. In an optimal situation, it is capable of excellent osteointegration with the bone and it is able to form a calcium phosphate-rich layer on its surface, Figure 4, very similar to hidroxyapatite which also prevents corrosion. Another advantageous property is that in case of damaging the protective layer the titanium oxides and calcium phosphate layer regenerate. Figure 4. Formation of hydroxyapatite layer on titanium oxide film[33] Slika 4. Nastanek hidroksiapatitne plasti na plasti titanovega oksida[33] The properties and biocompatibility of ni-tinol have their own characteristics which are different from those of nickel or titanium alone. In vitro NiTi biocompatibility studies on the effects of cellular tolerance and its cytotoxicity have been performed on various cell culture models[34,35]. Human monocytes and microvascular endothelial cells were exposed to pure nickel, pure titanium, stainless steel and nitinol. Nitinol has been shown to release higher concentrations of Ni2+ ions in human fibroblast and osteoblast cultures, which did not affect cell growth[36-38]. Metal ion release study also revealed very low concentrations of nickel and titanium that were released from nitinol. Researchers therefore concluded that nitinol is not genotoxic. For in vivo biocompatibility studies of nitinol effect, different experiments have been done on animals. Several in vivo ni-tinol biocompatibility studies which were done in the last decade disclosed no allergic reactions, no traces of alloy constituents in the surrounding tissue and no corrosion of implants. Studies of rat tibiae response to NiTi, compared with Ti-6Al-4V and AISI 316L stainless steel, showed that the number and area of bone contacts was low around NiTi implants, but the thickness of contact was equal to that of other implants. Normal new bone formation was seen in rats after 26 weeks after implantation. Good biocompatibility results of NiTi are attributed to the fact that implants are covered by a titanium oxide layer, where only small traces of nickel are being exposed. Corrosion Behavior The body is a complicated electrochemical system that constitutes an aggressive corrosion environment for implants which are surrounded by bodily fluids of an aerated solution containing 0.9 % NaCl, with minor amounts of other salts and organic compounds, serum ions, proteins and cells which all may modify the local corrosion effect. High acidity of certain bodily fluids is especially hostile for metallic implants. Acidity can increase locally in the area adjacent to an implant due to inflammatory response of surrounding tissues mediating hydrogen peroxide and reactive oxygen and nitrogen compounds. The local pH changes for infected tissues or near hae-matomas are relatively small, however these changes can alter biological processes and thereby the chemistry around the implant. It is known that small point corrosion or pitting prevails on surfaces of metallic implants. Another important feature is roughness of the surface which increases the reacting area of the implant and thereby add to total amount of corrosion. Therefore surface finishing is a major factor in improving corrosion resistance and consequently biocompatibility of medical devices[39- 40]. Corrosion resistance of SMA has also been studied in vivo on animals. Plates and stents have been implanted in dogs and sheeps for several months. Corrosion has been examined under microscope and pitting was established as predominant after the implants were removed. Thus surface treatments and coatings were introduced. The improvement of corrosion resistance was considerable, since pitting decreased in some cases from 100 pm to only 10 pm in diameter. Surface The human response to implanted materials is a property closely related to the implant surface conditions. The major problems associated with the implants currently used are inadequate implant-tissue interface properties. Parameters that characterize surface property are chemical composition, crystallinity and heterogeneity, roughness and wettability or surface free energy which is a parameter important for cell adhesion. Each parameter is of great importance to biological response of the tissue. Another problem is implant sterilization which can remarkably modify desired parameters. Steam and dry sterilization are nowadays replaced by more advanced techniques like hydrogen peroxide plasma, ethylene oxide, and electron and y-ray irradiation. The surface of NiTi SMA has revealed a tendency towards preferential oxidation of titanium. This behavior is in agreement with the fact that the free enthalpy of formation of titanium oxides is negative and exceeds in absolute value the enthalpy of formation of nickel oxides by at least two to three times. The result of oxidation is an oxide layer of a thickness between 2-20 nm, which consists mainly of titanium oxides TiO2, smaller amounts of elemental nickel Ni, and low concentrations of nickel oxides NiO. The surface chemistry and the amount of Ni may vary over a wide range, depending on the preparation method. The ratio of Ti/Ni on polished surface is around 5.5, while boiled or autoclaved items in water show decreased concentration of Ni on the surface and the Ti/Ni ratio increases up to 23 to 33 [41]. Different in vitro studies have shown how the physical, chemical and biocompatible properties of the implant surface can be improved[42-46]. Surface Improvements Some of the most important techniques for improving the properties of Ni-Ti alloy surfaces are: (1) Surface modification by using energy sources and chemical vapors like hydroxyapatite, laser and plasma treatment, ion implantation, TiN and TiCN chemical vapor deposits. Hydroxyapatite coatings result in the best known biocom-patibility and reveal a tendency to dissolution due to its relative miscibility with body fluids. Ion implantation and laser treatments usually result in surface amor-phization that improves corrosion resistance, but the obtained amorphous surface layers are often not uniform. Laser surface melting leads to an increased oxide layer, decrease of Ni dissolution and improvement of the cytocompatibility up to classical Ti level. There is also a possibility that laser melted surfaces may be enriched in nickel, and become harder than bulk and swell. TiN and TiCN coatings are known to improve corrosion resistance but large deformations caused by the shape memory effect may cause cracking of the coating. Therefore, for plates and staples a plasma-polymerized tetrafluoroethylene has been introduced. (2) Development of bioactive surfaces is another approach to improve biocompatibility of the SMA. Human plasma fibronectin covalently immobilised to NiTi surface improved the attachment of cells while corrosion rates were reduced drastically. Studies showed NiTi surface improved with this method caused a development of calcium phosphate layers, which in fact eliminate the need for hydroxyapatite coatings[43,47]. (3) Electrochemical processing for oxidation in air/ oxygen is a most common way of metal surface treatment. The technique combines electrochemical processes and oxidation in various media. Growth of native passive films that are highly adhesive and do not crack or break due to dynamic properties of SMA is promoted with this method. Oxide films obtained in air have different colors, thickness, and adhesive properties, with TiO2 as a predominant oxide type. (4) Oxidation of SMA medical devices in water and steam is also one of the surface improvement techniques. Implants are preliminary chemically etched and boiled in water. The result is a surface with a very low Ni concentration, while etching removes surface material that was exposed to processing procedures and acquired various surface defects and heterogeneity. It also selectively removes nickel and oxidizes titanium. Surfaces obtained after oxidation in steam show better properties than those oxidized in water. (5) Electrochemical techniques are commonly used to pas-sivate NiTi surfaces. Surface passivation using electropolishing is often considered as a treatment of first choice just because this technique is used for surface conditioning of stainless steels, Co-Cr alloys, etc. However, the universal techniques developed for surface passivation of various alloys used for medical purposes are not necessary efficient for NiTi. It should also be noted that the implant surface coatings are not always beneficial. The major problem of titanium based alloys is that the formation of TiO2, according to the chemical equation Ti + 2H2O ^ TiO2 + 4H+ + 4e-, reduces the pH level at the titanium/coating interface. This means that if the coating is composed of hydroxy-apatite, it can dissolve, which gradually leads to detachment of the coating. Medical applications The trends in modern medicine are to use less invasive surgery methods which are performed through small, leak tight portals into the body called trocars. Medical devices made from SMAs use a different physical approach and can pull together, dilate, constrict, push apart and have made difficult or problematic tasks in surgery quite feasible. Therefore unique properties of SMAs are utilized in a wide range of medical applications. Some of the devices used in various medical applications are listed below. Stents are most rapidly growing cardiovascular SMA cylindrical mesh tubes which are inserted into blood vessels to maintain the inner diameter of a blood vessel. The product has been developed in response to limitations of balloon angioplasty, which resulted in repeated blockages of the vessel in the same area. Ni-Ti alloys have also become the material of choice for super-elastic self-expanding (SE) stents which are used for a treatment of the superficial femoral artery disease, Figure 5(a). The SE nitinol stents are produced in the open state mainly with laser cut tubing and later compressed and inserted into the catheter. They can also be produced from wire and laser welded or coiled striped etched sheet. Before the compression stage, the surface of the stent is electrochemically polished and passivated to prescribed quality. Deployment of the SE stent is made with the catheter. During the operation procedure, when the catheter is in the correct position in the vessel, the SE stent is pushed out and then it expands against the inner diameter of the vessel due to a rise in temperature (thermally triggered device). This opens the iliac artery to aid in the normal flow of blood. The delivery catheter is then removed, leaving the stent within the patient's artery. Recent research has shown that implantation of a self-expanding stent provides better outcomes, for the time being, than balloon angioplasty[48-50]. The Simon Inferior Vena Cava (IVC) filter was the first SMA cardiovascular device. It is used for blood vessel interruption for preventing pulmonary embolism via placement in the vena cava. The Simon filter is filtering clots that travel inside bloodstream[51]. The device is made of SMA wire curved similary to an umbrella which traps the clots which are better dissolved in time by the bloodstream. For insertion, the device is exploiting the shape memory effect, i.e. the original form in the martensitic state is deformed and mounted into a catheter. When the device is released, the body's heat causes the filter to return to its predetermined shape. The Septal Occlusion System is indicated for use in patients with complex ventricular septal defects (VSD) of significant size to warrant closures that are considered to be at high risk for standard transatrial or transarterial surgical closure based on anatomical conditions and/or based on overall medical condition. The system consists of two primary components; a permanent implant, which is constructed of an SMA wire framework to which polyester fabric is attached, and a coaxial polyurethane catheter designed specifically to facilitate attachment, loading, delivery and deployment to the defect[52]. The implant is placed by advancing the delivery catheter through blood vessels to the site of the defect inside the heart. The implant remains in the heart and the delivery catheter is removed. Instruments for minimally invasive surgery used in endoscopic surgery could not be feasible without implementation of SMA materials. The most representative instruments such as guidewires, dilatators and retrieval baskets exploit good kink resistance of SMAs[53]. Open heart stabilizers are instruments similar to a steerable joint endoscopic camera. In order to perform baypass operations on the open heart stabilizers are used to prevent regional heart movements while performing surgery. Another employment of the unique properties of SMAs such as constant force and superelasticity in heart surgery is a tissue spreader used to spread fatty tissue of the heart, Figure 5(b). In general, conventional orthopedic implants by far exceed any other SMA implant in weight or volume. They are used as fracture fixation devices, which may or may not be removed and as joint replacement devices. Bone and nitinol have similar stress-strain characteristics, which makes nitinol a perfect material for production of bone fixation plates, nails and other trauma implants[54]. In traditional trauma surgery bone plates and nails fixated with screws are used for fixation of broken bones. Shape memory fixators are one step forward applying a necessary constant force to faster fracture healing. The SMA embracing fixator consists of a body and sawtooth arms[55]. It embraces the bone about 2/3 of the circumference, Figure 5(c). The free ends of the arms which exceed the semi-circle are bent more medially to match the requirement fixation of a long tubular body whose cross section is not a regular circle. The applied axial compression stress is beneficial for enhancing healing and reducing segmental osteoporosis caused by a stress shielding effect. Its martensitic transformation temperature is 4-7 °C and shape recovery temperature is around the body's normal temperature, 37 °C. Similar to the embracing fixator is the so called Swan-Like Memory-Compressive Connector (SMC) for treatment of fracture and nonunion of upper limb diaphysis. The working principle of the device is similar with one important improvement. The SMC trauma implant is able to put constant axial stress to a fractured bone[56]. For fixation of tibial and femoral fractures nails fixated with screws are normally used. New SMA inter-locking intramedullary nails have many advantages compared to traditional ones. For example, when cooled SMA interlocking nails are inserted into a cavity, guiding nails are extracted and body heat causes bending of nails into a preset shape applying constant pressure in the axial direction of the fractured bone[57]. The SMA effect is also used in surgical fixators made from wire. Certain device which have been developed to fix vertebra in spine fractures are similar to an ordinary staple. Staple shaped compression medical devices are also used for internal bone fixation[58]. The compression staple is one of most simple and broadly used SMA devices in medicine, Figure 5(d). Since its introduction in 1981, over a thousand patients have been all successfully treated using this device. The SMA Patellar Concentrator was designed to treat patellar fractures, Figure 5(e). The device exerts continuous compression for the fixation of patella fracture. The shape of patellar concentrator consists of two basic patellae claws, conjunctive waist and Figure 5. Examples of nitinol medical devices Slika 5. Primeri medicinskih pripomočkov iz nitinola three apex patellae claws. The thickness of the device may vary between 1.8 and 2.2 mm depending on different sizes of the concentrator. In clinical surgery, the claws are unfolded and put over fractured patella. Exposed to body temperature, the device tends to recover to its original state resulting in a recovery compressive force[59]. Dentists are using devices made from SMA for different purposes. NiTi based SMA material performs exceptionally at high strains in strain-controlled environments, such as exemplified with dental drills for root canal procedures. The advantage of these drills is that they can be bent to rather large strains and still accommodate the high cyclic rotations[52]. Superelastic SMA wires have found wide use as orthodontic wires as well, Figure 6(b). NiTi orthodontic archwire was first produced in batches and clinically used in China in the beginning of 1980's [60]. Due to its unique prop-erty-superelasticity, the wire exerts gentle and retentive force to teeth, which is superior to stainless steel wire. Shape memory bracelet do not require as frequent visits to the dentist as the classical ones because of their ability to self adjust. The therapeutic period is therefore cut down by 50 %. Lately a special fixator for mounting bridgework has been developed, Figure 6(a). A small piece of SMA metal is notch on both sides and placed between teeth and bridgework. As the temperature rises the notched area of metal is expanded on both sides causing a permanent hold of bridge-work. The tooth fixator can also be used to prevent a loose tooth from falling out. Conclusions A SMA implants and medical devices have been successful because they offer a possibility of performing less invasive surgeries. Nitinol wires in medical instruments are more kink resistant and have smaller diameters compared to stainless steel 316L or polymer devices. Research to develop composite materials, containing SMA which will prove cost efficient and porous SMAs which will enable the transport of body fluids through its bulk is currently underway. Figure 6. Dental applications of nitinol Slika 6. Zobozdravstvene aplikacije nitinola Povzetek References Materiali z oblikovnim spominom v medicini [1] Materiali z oblikovnim spominom so zgod- [2] ba o uspehu uporabe v medicini in še vedno pridobivajo na svoji uporabnosti. Velike prednosti s kirurškega stališča pomenijo ogromne možnosti za nove komercialne [3] aplikacije. To delo podaja pregled razvoja materialov z oblikovnim spominom, njihove fizikalne zakonitosti ter uporabnost v medicini. Predstavljeni so pojav oblikovnega spomina, psevdoelastičnost in druge osnovne lastnosti s spominom oblike. V nadaljevanju so predstavljeni medicinski pripomočki, ki izkoriščajo oblikovni spomin in sedanja komercialna uporaba. Ker je večina takšnih medicinskih pripočkov [5] izdelanih iz materiala nitinol so preučeni tudi vplivi, ki zadevajo nevarnosti uporabe tega materiala v medicinskih aplikacijah. Medicinski implantati in ostali pripomočki iz materialov z oblikovnim spominom so [7] uspešni predvsem zaradi zmožnosti izvajanja manj invanzivnih kirurških posegov. Žice in nitinola in medicinski instrumenti so bolj odporni proti prepogibom in imajo [8] v primerjavi z nerjavečim jeklom ali polimeri manjše prečne premere. V teku so številne raziskave, ki se ukvarjajo z razvojem in uporabnostjo kompozitnih materialov, ki v svoji sestavi vsebujejo tudi materiale s spominom oblike. Porozni materiali s spominom oblike, ki bi omogočali prehod telesnih sokov skozi prerez, pa so trenutno tik pred tem, da preidejo v uporabo za komercialne namene. [9] Olander, A. (1932): An electrochemical investigation of solid cadmium-gold alloys. J. Am. Chem. Soc.; Vol. 54, No. 3819-3833. Greninger, A.B., Mooradian, V.G. (1938): Strain transformation in metastable beta copper-zinc and beta copper-tin alloys. AIME.; Vol. 128, pp. 337-368. Chang, L.C., Read, T.A. (1951): Plastic deformation and diffusionless phase changes in metals-the gold-cadmium beta phase. Am. Inst. Min. Metall. Eng., J. Met.; Vol. 191/1, pp. 47-52. Buehler, W.J., Wang, F.E. (1967): A summary of recent research on the Niti-nol alloys and their potential application in ocean engineering. Journal of Ocean Engineering.; Vol. 1, pp. 105108. Wayman, C.M. (1964): Introduction to the crystallography of martensitic transformations. The Macmillan Company. Otsuka, K., Wayman, C.M. (1998): Shape memory materials. Cambridge University Press. Wechsler, M.S., Liberman, D.S., Read, T.A. (1953): On the theory of the formation of martensite. Trans. AIME.; Vol. 197, pp. 1503-1515. Bowles, J.S., Mackenzie, J.K. (1954): The crystallography of martensite transformations I. Acta Metallurgical Vol. 2, pp. 129-137. Saburi, T., Wayman, C.M. (1979): Crys-tallographic similarities in shape memory martensites. Acta Metallur-gica.; Vol. 27/6, pp. 979-995. [13] [19] [21] [22] [10] Adachi, K., Perkins, J., Wayman, c.M. [18] (1986): Type II twins in self-accommodating martensite plate variants in a Cu-Zn-Al shape memory alloy. Acta Metallurgical, Vol. 34/12, pp. 24712485. [11] James, R.D., Hane, K.F. (2000): Martensit- ic transformations and shape-memory materials. ActaMaterialia.; Vol. 48/1, pp. 197-222. [20] [121 Krishnan, M. (1998): The self accommodating martensitic microstructure of Ni-Ti shape memory alloys. Acta Materialia.; Vol. 46/4, pp. 1439-1457. Inamura, T., Kinoshita, Y., Kim, j.I., Kim, H.Y., Hosoda, H., Wakashima, K., Miyazaki, s. (2006): Effect of {0 0 1}<1 1 0> texture on superelastic strain of Ti-Nb-Al biomedical shape memory alloys. Materials Science and Engineering A.; Vol. 438-440, pp. 865-869. [14] Bhattacharya, K. (2003): Microstructure p3] of martensite: why it forms and how it gives rise to the shape-memory effect. Oxford Series on Materials Modelling, 1st Ed., Oxford University Press, [24] Oxford. [15] stalmans, R., Delaey, L., Van Humbeeck, J. (1997): Generation of recovery stresses: thermodynamic modelling and experimental verification. J. de Phys. IV; Vol. 7, pp. 47-52. [16] Barsch, G.R., Krumhansl, J.A. (1984): [25] Twin boundaries in ferroelastic media without interface dislocations. Phys. Rev. Lett.; Vol. 53/11, pp. 1069-1072. [17] Falk, F. (1980): Model free energy, me- [26] chanics, and thermodynamics of shape memory alloys. Acta Metall.; Vol. 28, pp. 1773-1780. Maugin, G.A., Cadet, S. (1991): Existence of solitary waves in martensitic alloys. Int. J. Eng. Sci. ; Vol. 29/2, pp. 243-258. Brinson, L.C., Lammering, R. (1993): Finite element analysis of the behavior of shape memory alloys and their applications,. Int. J. Solids and Struct.; Vol. 30/23, pp. 3261-3280. Ivshin, Y., Pence, T.J. (1993): A thermome-chanical model for a one variant shape memory material. J. Intell. Mat. Syst. and Struct.; Vol. 5/7, pp. 455-473. Liang, C., Rogers, C.A. (1990): One-dimensional thermomechanical constitutive relations for shape memory materials. J. Intell. Mater. Syst. and Struct.; Vol. 1/2, pp. 207-234. Boyd, J.G., Lagoudas, D.C. (1994): Ther-momechanical response of shape memory composites. J. Intell. Mater. Syst. and Struct.; Vol. 5, pp. 333-346. Tanaka, K. (1986): A thermomechanical sketch of shape memory effect: One-dimensional tensile behaviour. Res Mech.; Vol. 18, pp. 251-263. Brinson, L.C. (1993): One-dimensional constitutive behavior of shape memory alloys: thermomechanical derivation with non-constant material functions and redefined martensite internal variable. J. Intell. Mater. Syst. and Struct.; Vol. 4, pp. 229-242. Lubliner, J.F., Auricchio, F. (1996): Generalized plasticity and shape-memory alloys. Int. J. Solids and Structures.; Vol. 33/7, pp. 991-1003. Panoskaltsis, VP., Bahuguna, S., Sol-datos, D. (2004): On the thermomechanical modeling of shape memory alloys. Int. J. Non-Linear Mech. ; Vol. 39/5, pp. 709-722. [27] sun, Q.P., Hwang, K.c. (1994): Micro- mechanics constitutive description of thermoelastic martensitic transformations. Advances in Applied Mechanics.; Vol. 31, pp. 249-298. [28] Kosel, F., Videnic, T. (2007): Generalized plasticity and uniaxial constrained recovery in shape memory alloys. Mech. Adv. Mater. Struc.; Vol. 14/1, pp. 3-12. [29] Denkhaus, E., Salnikow, K. (2002): Nick- el essentiality, toxicity, and carcino-genicity. Critical Reviews in Oncol-ogy/Hematology.; Vol. 42, pp. 35-56. [30] Nieboer, E., Tom, R.T., Sanford, We. (1988): Nickel metabolism in man and animals (In: Nickel and Its Role in Biology: Metal Ions in Biological Systems), Vol. 23 (Sigel H, ed). New York: Marcel Dekker, pp. 91-121. [31] Fletcher, G.G., Rossetto, F.E., Turnbull, J.D., Nieboer, E. (1994): Toxicity, Uptake, and Mutagenicity of Particu-late and Soluble Nickel Compounds. Environ Health Perspect.; Vol. 102 (Suppl 3), pp. 69-79. [32] Yamamoto, A., Honma, R., Sumita, M. (1998): Cytotoxicity evaluation of 43 metal salts using murine fibroblasts and osteoblastic cells. Journal of Biomedical Materials Research.; Vol. 39, pp. 331-340. [33] combes, c., Rey, c., Freche, M. (1998): XPS and IR study of dicalcium phosphate dihydrate nucleation on titanium surfaces. Colloids and Surfaces B: Biointerfaces.; Vol. 11/1-2, pp. 1527. [34] shih, c., Lin, s., Chung, K., Chen, Y., su, Y., Lai, s., Wu, G., Kwok, c., chung, K. (2000): The cytotoxicity of corrosion products of Nitinol stent wires on cultured smooth muscle cells. J. Biomed. Mater. Res.; Vol. 52, pp. 395403. [35] Wever, D.J., Veldhuizen, A.G., sanders, M.M., schakenraad, j.M., Horn, J.R. (1997): Cytotoxic, allergic and geno-toxic activity of a nickel-titanium alloy. Biomaterials.; Vol. 18, pp. 11151120. [36] Wataha, I.c., Lockwood, P.E., Marek, M., Ghazi, M. (1999): Ability of Ni-containing biomedical alloys to activate monocytes and endothelial cells in vitro. J Biomed Mat Res.; Vol. 45, pp. 251-257. [37] Ryhänen, j., Niemi, e., serlo, W., Niemelä, e., sandvik, P., Pernu, h., salo, T. (1997): Biocompatibility of nickeltitanium shape memory metal and its corrosion behavior in human cell cultures. J Biomed Mat Res.; Vol. 35, pp. 451-457. [38] Wirth, c., comte, V., Lagneau, C., Ex- brayat, P., Lissac, M., JaFFREZIC-Renault, N., Ponsonnet, L. (2005): Nitinol surface roughness modulates in vitro cell response: a comparison between fibroblasts and osteoblasts. Materials Science and Engineering: C, Vol. 25, pp. 51-60. [39] Trepanier, C., Leung, T., Tabrizian, M., Yahia, L'H., Bienvenu, j., Tanguay, j., Piron, D., Bilodeau, L. (1999): Preliminary investigation of the effect of surface treatment on biological response to shape memory NiTi stents. J. Biomed. Mater. Res.; Vol. 48, pp. 165-171. [40] shabalovskaya, s.A. (2002): Surface, corrosion and biocompatibility aspects of Nitinol as an implant material. Biomed. Mater. Eng.; Vol. 12, pp. 69-109. [41] Shabalovskaya, s.A. (1996): On the na- ture of the biocompatibility and on medical applications of NiTi shape memory and superelastic alloys. Bi- omed.Mater.Eng.; Vol. 6, pp. 267289. [42] Frauchiger, VM., Schlottig, F., Gasser, B., Textor, M. (2004): Anodic plasma-chemical treatment of CP titanium surfaces for biomedical applications. Biomaterials.; Vol. 25, pp. 593-606. [43] Lu, x., zhao, z., Leng, Y. (2006): Biomi- metic calcium phosphate coatings on nitric-acid-treated titanium surfaces. Materials Science and Engineering: C.; Vol. 27/4, pp. 700-708. [44] Park, J., kim, D.J., kim, y.k., lee, K.H., Lee, H., Ahn, S. (2003): Improvement of the biocompatibility and mechanical properties of surgical tools with TiN coating by PACVD. Thin Solid Films.; Vol. 435/1-2, pp. 102-107. [45] Shevchenko, N., Pham, M.T., Maitz, M.F. (2004): Studies of surface modified NiTi alloy. Applied Surface Science.; Vol. 235, pp. 126-131. [46] Endo, K. (1995): Chemical modification of metallic implant surfaces with bio-functional proteins (Part 1). Molecular structure and biological activity of a modified NiTi alloy surface. Dent. Mater .J.; Vol. 14, pp. 185-198. [47] Liu, F., Wang, F., Shimizu, T., Igarashi, K., Zhao, L. (2006): Hydroxyapatite formation on oxide films containing Ca and P by hydrothermal treatment. Ceramics International.; Vol. 32/5, pp. 527-531. [48] Schillinger, M., Sabeti, S., Loewe, c. (2006): Balloon angioplasty versus implantation of nitinol stents in the superficial femoral artery. Journal of Vascular Surgery.; Vol. 44/3, pp. 684. [49] Rapp, B. (2004): Nitinol for stents. Materi- als Today.; Vol. 7/5, pp. 13. [50] Tyagi, S., Singh, S., Mukhopadhyay, S., Kaul, u.A. (2003): Self- and balloon-expandable stent implantation for severe native coarctation of aorta in adults. American Heart Journal., Vol. 146/5, pp. 920-928. [51] Simon, M., Kaplow, R., Salzman, E., Frei- man, D. (1977): A vena cava filter using thermal shape memory alloy Experimental aspects. Radiology.; Vol. 125, pp. 87-94. [52] Duerig, T., Pelton, A., Stockel, D. (1999): An overview of nitinol medical applications. Materials Science and Engineering: A.; Vol. 273-275, pp. 149160. [53] Fischer, H., Vogel, B., Grunhagen, A., Brhel, K.P., Kaiser, M. (2002): Applications of Shape-Memory Alloys in Medical Instruments. Materials Science Forum.; Vol. 394-395, pp. 9-16. [54] Pelton, A.R., Stockel, D., Duerig, T.W. (2000): Medical Uses of Nitinol. Materials Science Forum.; Vol. 327-328, pp. 63-70. [55] Dai, K., Wu, x., Zu, x. (2002): An In- vestigation of the Selective Stress-Shielding Effect of Shape-Memory Sawtooth-Arm Embracing Fixator. Materials Science Forum.; Vol. 394395, pp. 17-24. [56] zhang, c., xu, s., Wang, J., Yu, B., zhang, Q. (2002): Design and Clinical Applications of Swan-Like Memory-Com-pressive Connector for Upper-Limb Diaphysis. Materials Science Forum.; Vol. 394-395, pp. 33-36. [57] Da, G., Wang, T., Liu, Y. , Wang, c. (2002): Surgical Treatment of Tibial and Femoral Factures with TiNi Shape-Memory Alloy Interlocking Intramedullary Nails. Materials Science Forum.; Vol. 394-395, pp. 37-40. [58] song, c., Frank, T.G., campbell, P.A., cuschieri, A. (2002): Thermal Modelling of Shape-Memory Alloy Fixator for Minimal-Access Surgery. Materials Science Forum.; Vol. 394-395, pp. 53-56. [59] xu, s, Zhang, c., Li, s., su, j., Wang, j. (2002): Three-Dimensional Finite Element Analysis of Nitinol Patellar Concentrator. Materials Science Forum.; Vol. 394-395, pp. 45-48. [60] chu, y., dai, K., zhu, M., Mi, x. (2000): Medical Application of NiTi Shape Memory Alloy in China. Materials Science Forum.; Vol. 327-328, pp. 5562.