J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 309–319 MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS EXTERNAL EAR STIMULATION: A TWO-CASE STUDY MERITEV OSNOVNE HEMODINAMIKE MED VE^TO^KOVNO TRANSKUTANO STIMULACIJO ZUNANJEGA U[ESA: [TUDIJA NA DVEH OSEBAH Janez Rozman 1,2* , Katarina Miklavec 2 , Ingrid Rozman 3 , Anja Emri 4 , Tomislav Mirkovi} 5 , Samo Ribari~ 2 1 Center for Implantable Technology and Sensors, ITIS d. o. o. Ljubljana, Lepi pot 11, 1000 Ljubljana, Slovenia 2 Institute of Pathophysiology, Faculty of Medicine, University of Ljubljana, Zalo{ka 4, 1000 Ljubljana, Slovenia 3 The Family Study and Research Centre, Dvor 12, 1210 Ljubljana-[entvid, Slovenia 4 Medical Chamber of Slovenia, Dunajska cesta 162, 1000 Ljubljana, Slovenia 5 Department of Anaesthesiology and Surgical Intensive Therapy, University Medical Centre Ljubljana, Zalo{ka 2, 1000 Ljubljana, Slovenia Prejem rokopisa – received: 2023-10-26; sprejem za objavo – accepted for publication: 2024-04-04 doi:10.17222/mit.2023.1032 The objectives of the investigation were to assess the short-term responses of the cardio-vascular system on the multipost transcutaneous stimulation (tANS) of the external ear (EE). The scope was to measure the forefinger photopleths (FPPG), toe photopleths (TPPG), aortic phonocardiogram (APCG), mitral phonocardiogram (MPCG) and brachial arterial blood pressure (BABP), assuming that they could be altered with the tANS. For the tANS, stimulator, two silicone ear plugs with four platinum electrodes each, and a large common electrode (CE), were used. Trials were carried out with two female volunteers, aged 25 years and 28 years. To capture the heart sounds, two customized electronic stethoscopes (transducers) were used. BABP was measured using a pressure transducer and blood-pressure appliance. To measure the FPPGs, a pulse oximeter and SpO2 finger clip were used. To measure the TPPGs, another pulse oximeter and customized SpO2 foot clip was used. Signals were gathered using a high-performance data-acquisition system. An offline analysis was made just before and just after the start of the tANS. To record the pulsations in the right brachial artery, the 2D ultrasound mode (B) of the ultrasound device was used before and during the tANS. Then, diagrams showing the vein cross-sectional area over time were constructed. The results show that vascu- lar time intervals between heart sounds S1 and S2, captured as APCG and systolic peaks of the FPPG in the second volunteer, were slightly larger during the separate tANS of the left white (LW) and (RW) EE post than before the tANS. Furthermore, vas- cular time intervals between the heart sounds S1 and S2, captured as MPCG and systolic peaks of the TPPG in the second vol- unteer, were slightly smaller during the separate tANS of LW and RW EE post than before the tANS. Finally, tANS of the LW post elicited a slightly lower heart rate than the one measured when LW was not stimulated. In contrast, the tANS of the RW post elicited a slightly higher heart rate than the one measured when RW was not stimulated. It was also shown that the pattern of the vein cross-sectional area over time with tANS was different compared to both the FPPG and TPPG. Keywords: external ear stimulation, hemodynamics, measurement, signal acquisition Cilj raziskave je bil ovrednotiti kratkoro~ne odzive sr~no-`ilnega sistema na ve~to~kovno transkutano stimulacijo (tANS) zunanjega u{esa (EE). Raziskava je obsegala meritve fotopletismogramov na palcu desne roke (FPPG), fotopletismogramov na palcu leve noge (TPPG), aortnih fonokardiogramov (APCG), mitralnih fonokardiogramov (MPCG) in brahialnega arterijskaga krvnega tlaka (BABP) predpostavljajo~, da se bodo spremenili vsled tANS. Za tANS so bili uporabljeni stimulator, dva u{esna ~epka iz silikona s po {tirimi platinastimi ektrodami in velika skupna elektroda (CE). Meritve so bile opravljene na dveh prostovoljkah starih 25 let in 28 let. Za zajemanje zvokov srca sta bila uporabljena dva posebej izdelana elektronska stetoskopa (transducer). BABP pa je bil merjen s pretvornikom tlaka in merilcem krvnega tlaka. Za meritve FPPG je bil uporabljen pulzni oksimeter z naprstnim SpO2 senzorjem. Za meritve TPPG pa je bil uporabljen drugi pulzni oksimeter s posebnim SpO2 senzorjem na no`nem palcu. Signali so bili digitalizirani z zmogljivim sistemom za zajemanje podatkov. Kasnej{a analiza podatkov je bila narejena tik pred za~etkom in takoj po kon~ani tANS. Za bele`enje pulziranja desne brahijalne arterije je bil uporabljen 2D ultrazvo~ni na~in (B) ultrazvo~ne naprave in sicer pred in med tANS. Nato sta bila konstruirana diagrama, ki sta prikazovala presek arterije v odvisnosti od ~asa. Rezultati so pokazali, da so bili ~asovni intervali med zvoki srca S1 in S2 zajeti kot APCG in sistoli~nimi vrhi pri drugi prostovoljki, rahlo dalj{i tako med tANS levega belega (LW) kot tudi desnega (RW) mesta na EE, kot pred tANS. Nadalje so bili ~asovni intervali med zvoki srca S1 in S2 zajeti kot MPCG in sistoli~nimi vrhi pri isti prostovoljki, rahlo kraj{i tako med tANS levega belega (LW) kot tudi desnega (RW) mesta na EE, kot pred tANS. Razen navedenega je pri{lo v ~asu tANS mesta LW do rahlega zmanj{anja frekvenc sr~nega utripa kot je bil v ~asu brez tANS. Nasprotno pa je pri{lo v ~asu tANS mesta RW do rahlega pove~anja ritma srca kot je bil v ~asu brez tANS. Rezultati so tudi pokazali, da je bil ~asovni potek sprememb povr{ine prereza arterije v odvisnosti od ~asa med tANS razli~en tako od poteka FPPG kot tudi od poteka TPPG. Klju~ne beside: stimulacija zunanjega u{esa, hemodinamika, meritev, zajemanje signala Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 309 UDK 616.288:612.063 ISSN 1580-2949 Original scientific article/Izvirni znanstveni ~lanek MTAEC9, 58(3)309(2024) *Corresponding author's e-mail: jnzrzmn6@gmail.com (Janez Rozman) 1 INTRODUCTION Electrical nerve stimulation is the most often used technique to provide external control over body systems that are normally under the control of the nervous sys- tem. 1 Surgically implanted devices currently used for the stimulation of vagus nerve afferents have many disad- vantages. It was presumed, however, that most of them can be overcome using the novel method of trans- cutaneous vagus nerve stimulation tVNS. 2 Thus, tVNS devices that target the auricular (tANS) portion of the vagus nerve may have the potential to be implemented for rehabilitation, and treatment of physiological and mental disorders. 3 Namely, tANS technology targets afferents within the auricular branch of the vagus nerve that can be accessed from the external ear (EE). 4 Some studies reported effects of the tANS on the regulation of autonomic function through measurements of the heart rate, heart-rate variability, blood pressure, and related physiological quantities in healthy volunteers and in vas- cular hypertensive patients. 5 One non-invasive procedure used to describe the heart-function dynamics is phonocardiography (PCG). It is the graphical representation of heart sounds providing information useful to reveal abnormalities in the move- ment of the heart wall, closure of the valves, or the leak- age of blood flow. 6 In this regard, the PCG can be uti- lized to auscultate and distinguish four sounds during the heart cycle, i.e., S1, S2, S3, and S4, that arise from open- ing and closing of a valve. 7 Traditional positions of heart auscultation at the thorax are the following: the mitral area above the cardiac apex, the tricuspid area above the fourth and fifth intercostal space along the left sternal border, the pulmonic area above the second intercostal space along the left sternal border and aortic area above the second intercostal space along the right sternal bor- der. Heart sounds are complex and non-stationary subtle body sounds. They can be described by their intensity, pitch, position, quality and timing in the heart cycle. 8 Loudness of the heart is usually measured in decibels (dB) based on how far away a tested person is. To assess the heart function, a heart-sound activity detection framework has been developed by Varghees et al. 9 The authors captured PCG signals that included four heart sounds: S1, S2, S3 and S4. They also identified average heart sound durations, the time delay between heart sounds, and the duration of heart cycles. Finally, they identified normal frequency ranges of heart sounds (S1 (50–150) Hz, S2 (50–200) Hz, S3 (50–90) Hz, and S4 (50–80) Hz). Fast Fourier transform (FFT) was used for automatic heart-sound signal analysis. In healthy sub- jects, the two heart sounds, S1 and S2, are the most fre- quently used in the PCG analysis. 10 S1 results from the closing of the mitral and tricuspid valves. The sound pro- duced by the closure of the mitral valve is termed M1, and the sound produced by the closure of the tricuspid valve is termed T1. M1 is the main component of S1 and is much louder than T1. S2, however, is produced by the closure of the aortic and pulmonic valves where the clo- sure of the aortic valve is termed A2, and the closure of the pulmonic valve is termed P2. A2 is normally much louder than P2 and is the main component of S2. In gen- eral, heart valves have been considered passive structures that only function in response to the hemodynamic forces generated by heart contractions. 11 However, under physiological conditions, autonomic sympathetic nerves and the vagus nerve modulate the mechanical properties of heart valves. However, little focus has been given to the regulation of heart valves by the autonomic nervous system. 12 As the mitral valve is exposed to high mechani- cal forces during each heart cycle, it can be hypothesized that under such physiological conditions, mitral valve tissue stiffness varies in response to autonomic nervous agents and that the leaflet tone and valve function are modulated. Therefore, there is a need to identify the role of the autonomic nerve in the regulation of mitral valve mechanical properties. Oximeters used in pulse oximetry 13 often provide a patient’s heart rate via an assessment of the arterial blood pulsations. Pulse oximeters are also able to visualize a blood-volume change in the tissue caused by the passage of blood, and this is called a plethysmographic trace. It can resemble an arterial pressure waveform and is ap- proximated by the ratio of the stroke volume output to the compliance of the arterial tree. In clinical practice, photoplethysmography (PPG) is routinely used to moni- tor cardiac-induced blood volume changes at peripheral body sites, such as the finger, forehead, earlobe, and toe. 14 A normal PPG is characterized by a sharp systolic upstroke and peak and by a prominent dicrotic notch on the downward portion of the PPG. With compromised blood flow through an artery, however, the dicrotic notch can be lost or reshaped. A photoplethysmograph is often used to measure an optical PPG. It offers a simple and inexpensive method to detect blood-volume changes in the microvascular bed of tissue and to measure toe blood pressure. 15 In critically ill or hemodynamically unstable patients, however, there is a frequent need for hemodynamic mon- itoring such as cardiac output, oxygen saturation, and stroke volume variation. Hemodynamic monitoring is a procedure used in the repeated assessment of circulatory function to check blood circulation and to evaluate how well a heart is working over time. 16 The results of a hemodynamic test demonstrate how much blood a heart can pump and how well blood travels through the blood vessels. They can thus identify problems and suggest possibilities to address them. With this regard, the ultra- sound mode that is normally used to study vasculature is the two-dimensional (2-D) (B) mode. It uses the echo from solid structures to display 2-D images of the walls of blood vessels. 17 Data obtained by the most sophisti- cated duplex ultrasound instrument can show macro- vascular features of vascular physiology including imag- ing and flow velocity profiles. J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 310 Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 The present study was aimed at testing whether the multi post tANS of afferent nerve fibres within the auric- ular branch of the vagus nerve can be used as a method for the external modulation of basic hemodynamic func- tion. The main goal was to identify, record and analyse the short-term responses of the cardio-vascular system on the multipost tANS. The specific goal was to retain the characteristics of the main events within the mea- sured signals and to analyse them if they contained extractable information resulting from the tANS. With regard to the cardio-vascular system, the aim was to measure the forefinger photopleths (FPPG), toe photo- pleths (TPPG), PCG, and Brachial artery blood pressure (BABP), assuming that all can be altered with the tANS. 2 EXPERIMENTAL PART 2.1 Protocol Approval, Subject Consent and Subject Health Status This study was conducted on two female volunteers, aged 25 and 28 years. Anthropometric data for the first volunteer: age 25 years, height 174 cm, weight 63 kg, non-smoker, systolic blood pressure 115 mmHg and heart rate 60/min. Anthropometric data for the second volunteer: age 28 years, height 160 cm, weight 50 kg, non-smoker, systolic blood pressure 110 mmHg and heart rate 56/min. Both were in top physical condition and were free from any known cardio-vascular disease. They were recruited by asking them to participate in the study. Each participant was recruited for the study at dif- ferent times. We carried out the study approved by the National Medical Ethics Committee, Ministry of Health, Republic of Slovenia (No. 0120-297/2018/6). 2.2 Medical isolation and noise reduction The entire setup was connected to the human by con- sidering the class safety standard (IEC 60601: Interna- tional Product Safety Standards for Medical Devices). To accomplish the galvanical isolation between the mains supply section and that of the measuring setup section, the highest quality vintage 500 VA isolation transformer (supplied in 1966 with the STM-200-b studio Reel-to- Reel Mono Tape Recorder for professional use, Mechanikai Laboratórium (ML), Budapest, Hungary), was used. The isolation transformer used has a leakage current of 15 μA which was within the range of trans- formers rated for medical applications (between 10 and 50 μA). This power supply isolation also enabled a break of the ground loops and thereby the elimination of noise in the measuring setup. To additionally inhibit the noise transmitted through the grid power cable and high-fre- quency interference signals generated by the electronic device itself, the CE Certified Single-Phase 220V AC, 15A, EMI Power Supply Filter (EMI Filter) with Dual-Stage S Purification (CW4L2-I5A-S, Canny Well Co., Ltd., No. 333, Jen Chian St., Shu-Lin 23855, Tai- wan, China) was used. 2.3 Multipost transcutaneous stimulation set-up For the multi post tANS, the certified microproces- sor-controlled stimulator (Model SM9079, Shenzhen L-Domas Technology Ltd., Shenzhen, Guangdong, China) was used. The pattern of the stimuli train was se- lected from a palette of patterns available with the stimulator. The proprietary stimulating pulse was sym- metrical, current-regulated rectangular and dual direc- tional stimulating pulse pair. The parameters selected at the stimulator were the following: Frequency f = 45.5 Hz, Stimulating phase width tc = 200 μs, Anodic phase width ta = 200 μs, On time (pulse train duration) 3.32 s, Time gap between pulses train 2.8 s. To deliver the stimuli to a particular post at the EE, the negative output of the stimulator was connected to the corresponding electrode using a custom-developed switching unit. The common anode (CE), however, was connected to the positive output of the stimulator. For the multipost tANS, the two silicone ear plugs, having four platinum electrodes each, were developed. They were mounted onto the frame of the headphones and inserted J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 311 Figure 1: Stimulating system: a) stimulating plug, b) frame of head- phones with stimulating plugs, c) electrical stimulator, d) switching unit, e) common anode into the EE such that each cathode was in contact with the predefined post at the EE. The reusable CE was crafted using a 2-mm-thick, 300-mm-long and 25-mm-wide ribbon made of highly water-absorptive sponge that was stitched below the stainless-steel mesh and Velcro tape. The geometric surface of the anode was approximately 7500 mm 2 . When wetted, the approxi- mately 3-mm-thick anode contacted the skin as a homo- geneous conductive solution. The components of the stimulating system are shown in Figure 1. Despite the fact that the tANS was delivered via a battery-powered stimulator, tANS generated relatively large electric fields that propagated through the body to- wards the sensors of the measuring set-up. Fortunately, the sensors applied were out of the galvanic contact with the skin so the signals recorded could not be contami- nated by electrical artefacts. To eliminate any potential interference, the neck was selected as the location to at- tach the CE and the subject grounded. Thus, the location was away from the sensors, and it separated galvanically the head from the body via thick bone and other tissues at a relatively small cross-section of the neck. 2.4 Measuring devices and procedures To build the multi-channel measuring set-up, various custom-designed and commercially available devices were used. To capture heart sounds (PCG), two cus- tom-developed transducers were used. Brachial arterial blood pressure (BABP) was measured using a pressure transducer and commercial blood-pressure appliance. To measure the forefinger pleths (FPPG), a pulse oximeter and an adult SpO2 finger clip were used. To measure the toe pleths (TPPG), a pulse oximeter and customized photo-plethysmography foot clip were used. To record the blood flow in the brachial artery, the most advanced portable diagnostic ultrasound device was used. Signals were gathered using a high-performance I/O data-acqui- sition system and portable computer. 2.4.1 Auscultation of heart sounds To capture the PCG that came from the heart valves, two custom-developed transducers shown in Figure 2a were used (frequency range 20–20,000 Hz, analogue voltage output 0.2–1 V). To distinguish types of sounds namely, S1 and S2, 18 two transducers were placed at the standard heart auscultation positions shown in Figure 3. The first transducer was placed at the aortic heart auscultation position located at the 2 nd right intercostal space of the ribs near the sternum. The second transducer was placed at the mitral or apex heart auscultation posi- tion located at the 5 th left intercostal space of the ribs and in the middle of the clavicular line. To reduce the effect of high-frequency noises, the PCG signals were filtered using a low-pass filter (LPF) Butterworth, 10–300 Hz, 3 rd with cut-off frequency of 300 Hz. 19,20 The transducers were used to potentially differentiate types of particular valve sounds that may be modified with the tANS. For the calibration of both transducers, a portable sound- pressure calibrator (AWA6221A, Hangzhou Aihua In- struments Co., Ltd., Zhejiang, China) which provided a normal sound pressure level of 94 dB was used. The gain of each transducer was then trimmed until the same value as the normal sound-pressure level was obtained. The transducers were calibrated at room conditions of approximately 15 dB. For the frequency-response tests, however, a sine tone sweeping between 20 Hz and 1000 Hz was delivered to the transducers and the response was measured using the digital oscilloscope (InstruStar ISDS205A, Harbin ViMu Electronic Technology Co., Ltd., Harbin, China). To accomplish a FFT, the win- dow-weighting function and range of waveform data that provided the best results in signal processing were se- lected. During the recording of heart sounds, the low-pass Butterworth filter at 1000 Hz was selected. 2.4.2 Measurement of brachial arterial blood pressure The BABP was measured using a pressure transducer (Type 4-327-C, Range: 0–400 mmHg, Beckman Instru- ments Inc., Fullerton, CA, U.S.A.) shown in Figure 2b, that was situated serially between the proprietary cuff and commercial blood-pressure appliance (Oberarm- Blutdruckmessgerät MD 12450, salvatec, Essen, Ger- many). The appliance corresponds to standards for non-invasive blood-pressure measuring devices (EN 1060-1:1995+A1:2002 and EN 1060-3:1997+A1:2005). The pressure transducer and the appliance were cali- brated occasionally using the reference values provided by a digital non-invasive sphygmomanometer calibrator (SLK-BXY-250, Shelok, Shaanxi, China). In an off-line analysis of the BABP trajectory, the value of the systolic blood pressure was taken from the trajectory at the point in which the first BABP oscillation was detected. The value of the diastolic pressure, however, was taken from the trajectory at the point at which the oscillation starts to disappear. 2.4.3 Measurement of forefinger photopleths To measure the FPPG, a pulse oximeter (Nellcor N-595, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA, U.S.A.) and an adult SpO2 finger clip attached on the right forefinger (Nellcor DS-100A, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA, U.S.A.) shown in Fig- ure 2c were used. 2.4.4 Measurement of toe photopleths To measure the TPPG, a pulse oximeter (Nellcor N-600, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA, U.S.A.) instrumented with the customized photo-plethysmography foot clip shown in the Figure 2d was used. For this purpose, an IR/LED pair and photosensor were removed from the an- imal ear clip SpO2 sensor (CSL032H, Oximax Animal Ear Clip SpO2 Sensor for Nellcor N-595, Shenzhen YKD Technology Co., Ltd., Guangdong, China), was at- J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 312 Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 tached to the two halves of the tube made of carbon fibre so to face the lower and upper sides of the toe, respec- tively. By doing so, a photosensor of the toe clip was ef- fectively shielded from direct light and the optical path from IR/LED pair to the photosensor was in an almost straight line. For the measurements, the foot clip was at- tached around the left toe. 2.4.5 Measurement of pulsations in brachial artery To record the blood flow in the brachial artery of the right arm in the first volunteer, two-dimensional (2-D) ultrasound mode (B), which uses the echo from solid structures to display 2-D images of the walls of blood vessels, was accomplished. For these purposes, the most advanced portable diagnostic ultrasound device (Venue Go™ Point of Care Ultrasound, ©GENERAL ELEC- TRIC COMPANY, U.S.A.), was used before and during the tANS. The transducer used was a linear array, which produced an array of parallel, high-frequency ultrasound while preserving detailed structural definition to a depth for the superficial brachial artery. The transducer was po- sitioned along the brachial artery using water-soluble ul- trasound transmission gel (Ultragel, AquaUltra Basic, Ultragel Hungary 2000 Ltd., Budapest, Hungary) and held in place. The position of the transducer was ad- justed until the clearest picture was obtained. The instru- ment was then switched automatically to pulse wave Doppler mode and the mean velocity and cross-sectional luminal area were measured simultaneously. Accord- ingly, the diameter of arteries that varied with blood flow status and over the cardiac cycle, was displayed and re- corded with motion. Records were then opened using the VLC media player free software and processed using ImageJ-win64 software. Processing was performed within three steps: Extraction of frames at a specific in- terval opened with VLC software, Image analysis and Data processing. Results were used to construct dia- grams showing a train of blood pulses expressed as vein cross-sectional area over time. 2.4.6 Signal acquisition All signals obtained from the conditioning circuits were gathered at 20 kHz with 24-bit resolution using a high-performance I/O data-acquisition system (DAC) (DEWE-43a, DEWESOFT d. o. o., Republic of Slo- venia) as shown schematically in Figure 3. A DAC has eight differential analogue channel inputs and a USB 2.0 interface. The stimulating intensity i c was assessed con- tinuously by measurement of a voltage drop across the precision serial resistor of 10Ù in the switching module connected to the stimulator output. During the acquisi- tion, each signal was filtered using the low-pass filter that was selected based on the frequency band of the par- ticular signal (see legend of Figure 3). Finally, data were stored on a portable computer (Lenovo W541, Lenovo, Beijing, China) to permit subsequent frequency analysis using software (DewesoftX). Figure 3 shows a sche- matic diagram of the components comprising the system while the legend below Figure 3 depicts the measured quantities and filtering applied. 2.4.7 Experimental Procedures A 45-minute trial was divided into four 5-minute se- quences where i C was delivered separately onto each of the four posts either at the left or the right EE, respec- J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 313 Figure 2: Measuring setup: a) Transducers for auscultation of phono-cardiographic signal, b) BABP pressure sensor, c) Forefinger clip SpO2 sen- sor, d) Customized toe SpO2 sensor tively. The four sequences were separated with 5-minute shams where the tANS was not delivered onto any of po- sitions. The 45-minute trial also started and ended with the 5-minute sham segment. Quantities i c , APCG, MPCG, FPPG, and TPPG were acquired continuously throughout the entire trial. The BABP, however, was ac- quired approximately at the middle of each four 5-min- ute sequence. The trials were carried out under the same (and as steady as possible) conditions with two healthy female volunteers, age 25 and 28. The subjects were instructed to avoid stress/tense and physical activity before the trial. Consumption of alcohol was prohibited before the trial. Consumption of a large meal later than an hour before the trial was prohibited. The subjects were asked to lay supine on the pedicure chair with arms at about heart level. The subjects were asked not to talk and to remain as still as possible. Stimulating and all testing positions were degreased with 70 % isopropyl alcohol and allowed to dry. The subjects were instructed to relax and rest their elbows while the BABP cuff was wrapped around the left arm. Washers were adhered to the transducers. Transducers were placed at the two standard heart auscultation posts. A SpO2 clip sensor was attached to the right forefinger and placed on the left toe. Head- phones with stimulating plugs were placed onto the head. The CA was placed on the neck. The intensity i c of the stimuli delivered to the posts LW or RV was set at the level just below when discomfort was detected. Muscles in the vicinity of the brachial artery were relaxed so the mechanical limitation of instantaneous brachial arterial flow was avoided. The experiment was ended while min- imizing i c and turning the stimulator off. The subjects re- mained lying supine so stimulating and sensory compo- nents were removed. The skin at the EE posts and the skin at the cervical neck was degreased and dried. 2.4.8 Offline Signal Analysis Offline signal analysis was carried out using a porta- ble computer (Lenovo W541, Lenovo, Beijing, China) and the aforementioned software DewesoftX. Any peri- ods containing motion artefacts that could not be consid- ered as captured signals were deleted. It was presumed that some potentially useful information about the state of the heart valves and arteries during and after tANS, would likely be obtained from the analysis. The BABP, FPPG, APCG, MPCG and TPPG, were analysed from traces during the BABP timeframe just before and just after the start of the tANS. 3 RESULTS The results of the assessment are depicted in Figure 4, Figure 5 and Figure 6. Figure 4 depicts the relation- J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 314 Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 Figure 3: Schematic diagram of the experiment Legend Mark Devices and conditions Input Measured Signal Symbol Unit Low Pass/Order a Stimulating posts R, Y, B, W N/A N/A N/A post N/A b Stimulating post selector N/A N/A N/A selector N/A c Common anode N/A N/A N/A anode N/A d Electrical stimulator 0 tANS intensity ic mA Bessel, 1 kHz, 4th e Personal computer N/A N/A PC N/A N/A f Auscultation transducer at aortic position 1 Aortic phono-cardiographic APCG dB Butterworth, 10–300 Hz, 4th g Auscultation transducer at mitral position 4 Mitral phono-cardiographic MPCG dB Butterworth, 10–300 Hz, 4th h Acquisition system N/A N/A DAC N/A N/A i Pressure transducer at blood pressure appliance 7 Brachial artery blood pres- sure BABP mmHg Bessel, 10 Hz, 4th j Power Supply Filter N/A N/A N/A N/A N/A k Pulse oximeter I 9 Forefinger photo-plethysmographic FPPG mmHg Bessel, 10Hz, 4th l Pulse oximeter II 14 Toe photo-plethysmographic TPPG mmHg Bessel, 10Hz, 4th m Isolation transformer N/A N/A N/A N/A N/A ship between FPPG, TPPG, BABP, APCG and MPCG, before and during the tANS. 9,21 Precisely, the results are represented through waveforms of the quantities APCG, MPCG, BABP, FPPG and TPPG that were recorded in the second volunteer before and during the tANS of the post LW located at the bottom of the left EE. Table 1 shows vascular time interval VTT1 elapsed between the first heart sound S1 and vascular time interval VTT2 elapsed between the second heart sound S2 and systolic peak in the FPPG signal before the tANS, vascular time interval VTT3 elapsed between the first heart sound S1 and vascular time interval VTT4 elapsed between the second heart sound S2 and systolic peak in the FPPG signal during the tANS. Table I also shows a vascular time interval VTT1 elapsed between the first heart sound S1 and vascular time interval VTT2 elapsed between the second heart sound S2 and systolic peak in the TPPG signal before the tANS, vascular time interval VTT3 elapsed between the first heart sound S1 and vascular time interval VTT4 elapsed between the second heart sound S2 and systolic peak in the TPPG signal during the tANS. Similarly, Figure 5 depicts the relationship between FPPG, TPPG, BABP, APCG and MPCG, before and dur- ing the tANS. 9 The results are represented through wave- forms of the quantities APCG, MPCG, BABP, FPPG, and TPPG that were recorded in the second volunteer be- fore and during the tANS of the post RW located at the bottom of the right EE. Table 1 shows vascular time in- terval VTT1 elapsed between the first heart sound S1 and vascular time interval VTT2 elapsed between the second heart sound S2 and systolic peak in the FPPG signal before the tANS, vascular time interval VTT3 elapsed between the first heart sound S1 and vascular time interval VTT4 elapsed between the second heart sound S2 and systolic peak in the FPPG signal during the tANS. Table I also shows a vascular time interval VTT1 elapsed between the first heart sound S1 and vascular time interval VTT2 elapsed between the second heart sound S2 and systolic peak in the TPPG signal before the tANS, vascular time interval VTT3 elapsed between the first heart sound S1 and vascular time interval VTT4 elapsed between the second heart sound S2 and systolic peak in the TPPG signal during the tANS. Figure 6 shows an ultrasound image of pulsing the right brachial artery indicated with arrows, and a train of pulsing vein cross-sectional area over time in the first volunteer. Figure 6a shows the brachial artery pulsing without tANS, while Figure 6b shows brachial artery pulsing with tANS of the W post at the left EE. Finally, J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 315 Figure 4: Relationship between APCG and MPCG, FPPG and TPPG, vascular time interval VTT1 between the first heart sound S1, vascular time interval VTT2 between the second heart sound S2, and the corresponding systolic peak in the FPPG and TPPG signals before and during multipost tANS of the LW in the second volunteer. Recorded quantities: a) i c, (b) APCG, c) MPCG, d) BABP, s) FPPG, and f) TPPG. J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 316 Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 Figure 5: Relationship between APCG and MPCG, FPPG and TPPG, vascular time interval (VTT1) between the first heart sound S1, vascular time interval (VTT2) between the second heart sound S2, and the corresponding systolic peak in the FPPG and TPPG signals before and during multipost tANS of the RW in the second volunteer. Recorded quantities: a) i c, b) APCG, c) MPCG, d) BABP, s) FPPG, and f) TPPG. Table 1: Vascular time intervals between heart sounds and systolic peaks, heart-sound amplitudes and heart-cycle durations in the second volun- teer before and during the tANS, respectively. Figure 6c shows the encircled vein as an important pro- cessing step of both video recordings. 4 DISCUSSION In this study we evaluated the effects of multipost tANS on the dynamics in heart activity and blood flow in two volunteers during tANS at predefined posts of the EE and listed potential effects on related hemodynamic functions. The overall hypothesis of the study was that the tANS of predefined posts on the left and right EE can have a measurable effect on heart sounds and some hemodynamic functions in healthy female volunteers. The main difference between our study and the stud- ies of others is that in our study the tANS can be deliv- ered to posts on the EE via multiple combinations of four electrodes mounted at predefined positions on a silicone plug. 22 In addition, optimum working electrochemical conditions of platinum cathodes used for the tANS were well defined considering Faradaic reactions as shown in Section 2.3 tANS set-up. As a result, skin irritation at the tANS posts was never observed. Furthermore, the charge density injected by the CE was lower than that injected by the cathode, so skin irritation stimulating effects were never observed. The greatest weakness of the paradigm, however, was that i c density and tANS efficiency were both dependent predominantly on the pressure produced by the silicone plug and less on small changes in the position of the plug within the EE. 23 Furthermore, trials were carried out and analysed by the same investigator, which might have re- sulted in some biases. However, the numerical data of the captured signals based on offline signal analysis shown in Table 1, minimized subjective interpretation. The study was intended mainly to obtain initial informa- tion on the efficiency of the multipost tANS so the num- ber of subjects tested and measurements performed were low. Accordingly, the results obtained should be consid- ered as a basis to gain further research activities that may result in clinically valuable conclusions. A discussion on the particular entity measured is presented below. Body sounds contain important information about human physiological and psychological conditions. 24 A non-invasive capture of biomedical signals is relatively simple and inexpensive; however, body sounds are usu- ally barely audible. Most contact transducers available are capable of capturing body sounds that are mainly lo- cated within the frequency spectrum, ranging from 20 Hz to 1.3 kHz. 25 Besides, the loudness of captured heart sounds varies with the auscultation positions. 26 However, modern biomedical signal-processing techniques are able to accurately characterize significant features of the heart sounds contained within the PCGs. 27 With this relation, the captured S1 and S2 sound can be used to determine the heart sound type and to detect potentially modified heart sounds. The amplitude of S1 may provide some valuable information about myocardial contractility abil- ity. 28 It was shown that transducers developed enabled a high-quality auscultation of S1 and S2. The hypothesis that the tANS has a measurable effect on heart sounds was confirmed. An artery of interest for the work was the brachial ar- tery as the major blood vessel of the upper arm. It was presumed that tANS can elicit a measurable difference in the BABP pressure. This hypothesis was not confirmed. Results showed that vascular time intervals VVT1, VVT2, VVT3 and VVT4 between heart sounds S1 and S2, captured as APCG and systolic peaks measured as FPPG in the second volunteer, were slightly larger dur- ing the separate tANS of both posts, once for the LW and once for the RW than before the tANS. It was also shown that during the tANS of the LW post, heart cycles S1-S1 and S2-S2 contained within the APCG, became slightly larger and, thus, the heart rhythm became slightly lower. Besides, they both became slightly louder. During the tANS of the RW post, however, heart cycles S1-S1 and S2-S2 contained within the APCG, became slightly smaller and, thus, the heart rhythm became slightly higher. Furthermore, S1 became slightly quieter while S2 became slightly louder. The hypothesis that tANS of the LW and RW has a measurable effect on the forefinger photopleths was confirmed. Results showed that vascular time intervals VVT1", VVT2", VVT3" and VVT4" between heart sounds S1 and S2, captured as MPCG and systolic peaks measured as TPPG in the second volunteer, were slightly smaller during the separate tANS of both posts, once for the LW and once for the RW than before the tANS. J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 317 Figure 6: Ultrasound image of the right Brachial artery pulsing and vein cross-sectional area over time in the first volunteer: a) without tANS, b) with tANS of the RW post, c) Phase of processing It was also shown that during the tANS of the LW post, heart cycles S1-S1 and S2-S2 contained within the MPCG became slightly larger and, thus, the heart rhythm slightly lower. Besides, both heart sounds S1 and S2 be- came louder. During the tANS of the RW post, however, heart cycles S1-S1 and S2-S2 contained within the MPCG, became slightly smaller and, thus, the heart rhythm became slightly higher. Furthermore, both, S1 and S2 became slightly quieter. The hypothesis that tANS of the LW and RW has a measurable effect on the toe photopleths was confirmed. The paper details how changes in pulsations of the brachial artery potentially elicited with the tANS were obtained using an ultrasound transducer. The exact dy- namics between the brachial artery pulsations and blood pressure are actually complex. Namely, blood vessels have viscoelastic properties that allow the diameter to vary with a pulsating pressure generated by the left ven- tricle contractions. It was shown in Figure 6a that the pattern of the right brachial artery pulsing expressed with a train of pulsing vein cross-sectional area over time without tANS, was different to the brachial artery puls- ing with tANS of the W post at the left EE. It was also shown that the pattern of the proprietary vein cross-sec- tional area over time with tANS was different compared to both the FPPG and TPPG. Accordingly, the hypothe- sis that tANS has a measurable effect on pulsations in the brachial artery was confirmed. Previous studies have indicated that the performance and autonomic function could be enhanced with the tANS and that tANS may be a promising treatment for some neuropsychiatric disorders. 29,30 With this relation the directions of our future work can include tANS using a matrix of electrodes and multiple combinations of elec- trodes to stimulate multiple posts on the EE and fine-tun- ing of the tANS parameters. 22 5 CONCLUSIONS It was shown that differences in all vascular time in- tervals and heart-cycle durations, obtained without tANS and with tANS of both posts, the LW and RW, were actu- ally small but were consistent. It can be concluded that tANS of the LW post elicited slightly lower heart rate than the one measured when LW was not stimulated. In contrast, it can be concluded that the tANS of the RW post elicited a slightly higher heart rate than the one measured when RW was not stimulated. Accordingly, the research involving tANS can pro- mote the method potentially useful to re-establish imbal- ance or lack of autonomic nervous modulation of heart valve function. One particular challenge arose from the study was to determine if valvar motion during diastole and/or systole is influenced by the innervation that was modulated by the multipost tANS. However, to place a methodology into service in a clinical environment, it should be tested in a larger group of subjects and trials. Acknowledgement The authors wish to acknowledge the Program team (Research core funding No. P3-0171 by the Slovenian Research and Innovation agency, Ministry of Higher Ed- ucation, Science and Innovation, Republic of Slovenia) at the Institute of Pathophysiology, Faculty of Medicine, University of Ljubljana, Republic of Slovenia for provid- ing language help, writing assistance, proof reading the article and editing the manuscript. 6 REFERENCES 1 J. T. Mortimer, N. Bhadra, Fundamentals of Electrical Stimulation, In: Krames ES, Peckham PH, Rezai Ali R, editors, Neuromodu- lation, Academic Press; 2009, 109–121. doi:10.1016/B978-0-12- 374248-3.00012-4 2 J. Ellrich, Transcutaneous vagus nerve stimulation, Eur. Neurol. Rev .,6( 2011) 4, 254–6, doi:10.17925/ENR.2011.06.04.254 3 W-P, Teo, M. Muthalib, S. Yamin, A. M. Hendy, K. Bramstedt, E. Kotsopoulos, S. Perrey, H. Ayaz, Does a combination of virtual real- ity, neuromodulation and neuroimaging provide a comprehensive platform for neurorehabilitation? A narrative review of the literature, Frontiers in Human Neuroscience, 10 (2016) 284, doi:10.3389/ fnhum.2016.00284 4 B. W. Badran, L. T. Dowdle, O. J. Mithoefer, N. T. LaBate, J. Coats- worth, J. C. Brown, W. H. DeVries, C. W. Austelle, L. M. McTeague, M. S. George, Neurophysiologic effects of transcutaneous auricular vagus nerve stimulation (taVNS) via electrical stimulation of the tragus: A concurrent taVNS/fMRI study and review, Brain Stimul., 11 (2018) 3, 492–500, doi:10.1016/j.brs.2017.12.009 5 B. W. Badran, O. J. Mithoefer, C. E. Summer, N. T. LaBate, C. E. Glusman, A. W. Badran, W. H. DeVries, P. M. Summers, C. W. Austelle, L. M. McTeague, J. J. Borckardt, M. S. George, Short trains of transcutaneous auricular vagus nerve stimulation (taVNS) have parameter-specific effects on heart rate, Brain Stimul., 11 (2018) 4, 699–708, doi:10.1016/j.brs.2018.04.004 6 M. A. A. Hamid, M. Abdullah, N. A. Khan, Y. M. A. AL-Zoom, Biotechnical System for Recording Phonocardiography, International Journal of Advanced Computer Science and Applications (IJACSA), 10 (2019) 8, doi:10.14569/IJACSA.2019.0100864 7 A. Ramovi}, L. Bandi}, J. Kevri}, E. Germovi}, A. Subasi, Wavelet and Teager energy operator (TEO) for heart sound processing and identification, In: CMBEBIH 2017, Springer; 2017, 495-502. https://www.healio.com/cardiology/learn-the-heart/cardiology-re- view/topic-reviews/heart-sounds 8 Learn The Heart. Cardiology Review, Topic Reviews A-Z, Heart Sounds Topic Review, https://www.healio.com/cardiology/learn-the- heart/cardiology-review/topic-reviews/heart-sounds, 20.06.2023 9 V. N. Varghees, K. I. Ramachandran, A novel heart sound activity detection framework for automated heart sound analysis, Biomedical Signal Processing and Control, 13 (2014), 174–188, doi:10.1016/ j.bspc.2014.05.002 10 T. H. Chowdhury, K. N. Poudel, Y. Hu, Time-Frequency Analysis, Denoising, Compression, Segmentation, and Classification of PCG Signals, IEEE Access, 8 (2020) 160882–160890, doi:10.1109/AC- CESS.2020.3020806 11 Y. Henderson, F. E. Johnson, Two modes of closure of the heart valves, Heart, 4 (1912), 69–82 J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... 318 Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 12 X. Hu, Q. Zhao, X. Ye, Autonomic regulation of mechanical proper- ties in porcine mitral valve cusps, Arq. Bras. Cardiol., 98 (2012) 4, 321–8, doi:10.1590/s0066-782x2012000400006. PMID: 22735910 13 A. Jubran, Pulse oximetry, Crit Care, 19 (2015) 1, 272, doi:10.1186/ s13054-015-0984-8 14 M. Elgendi, R. Fletcher, Y. Liang, N. Howard, N. H. Lovell, D. Abbott, K. Lim, R. Ward, The use of photoplethysmography for as- sessing hypertension, npj Digit. Med. 2 (2019) 60, doi:10.1038/ s41746-019-0136-7 15 P. Bonham, Measuring toe pressures using a portable photoplethysmograph to detect arterial disease in high-risk patients: an overview of the literature, Ostomy Wound Manage, 57 (2011) 11, 36–44 16 Cleveland Clinic. Hemodynamic Test, https://my.clevelandclinic.org/ health/diagnostics/17094-hemodynamic-test, 20.06.2023 17 J. P. Jamison, A. Campbell, C. Devlin, C. D. Johnson, Brachial artery blood flow by vascular ultrasound in education, Adv. Physiol. Educ., 46 (2022), 498–506, doi:10.1152/advan.00157.2021 18 A. Bhaskar, A simple electronic stethoscope for recording and play- back of heart sounds, Adv. Physiol. Educ., 36 (2012), 360–362, doi:10.1152/advan.00073.2012 19 A. M. Noor, M. F. Shadi, The heart auscultation. From sound to graphical, Journal of Engineering and Technology (JET), 4 (2013) 2, 73–84 20 C. N. Gupta, R. Palaniappan, S. Rajan, S. Swaminathan, S. M. Krishnan, Segmentation and classification of heart sounds, Proc. of the Can. Conf. Electrical Comput. Eng., Saskatoon 2005, 1674–1677, doi:10.1109/CCECE.2005.1557305 21 M. Nitzan, B. Khanokh, Y. Slovik, The difference in pulse transit time to the toe and finger measured by photoplethysmography, Physiol. Meas., 23 (2002) 1, 85–93. doi:10.1088/0967-3334/23/ 1/308 22 B. W. Badran, J. C. Brown, L. T. Dowdle, O. J. Mithoefer, N. T. LaBate, J. Coatsworth, W. H. DeVries, C. W. Austelle, L. M. McTeague, A. Yu, M. Bikson, D. D. Jenkins, M. S. George, Tragus or cymba conchae? Investigating the anatomical foundation of transcutaneous auricular vagus nerve stimulation (taVNS), Brain Stimul., 11 (2018) 4, 947–948, doi:10.1016/j.brs.2018.06.003 23 B. W. Badran, A. B. Yu, D. Adair, G. Mappin, W. H. DeVries, D. D. Jenkins, M. S. George, M. Bikson, Laboratory Administration of Transcutaneous Auricular Vagus Nerve Stimulation (taVNS): Tech- nique, Targeting, and Considerations, J. Vis. Exp., 143 (2019), e58984, doi:10.3791/58984 24 M. Elgendi, TERMA Framework for Biomedical Signal Analysis: An Economic-Inspired Approach, Biosensors, 55 (2016) 6(4), doi:10.3390/bios6040055 25 T. Rahman, A. T. Adams, M. Zhang, E. Cherry, B. Zhou, H. Peng, T. Choudhury, BodyBeat: a mobile system for sensing nonspeech body sounds, Proc. of the 12th annual international conference on Mobile systems, applications, and services, Bretton Woods 2014, 2–13, doi:10.1145/2594368.2594386 26 M. McGregor, M. B. Rappaport, H. B. Sprague, A. L. Friedlich, The calibration of heart sound intensity, Circulation, 13 (1956) 2, 252–256., doi:10.1161/01.cir.13.2.252 27 W. C. Kao, C. C. Wei, Automatic phonocardiograph signal analysis for detecting heart valve disorders, Expert Systems with Applica- tions, 38 (2011) 6, 6458–6468, doi:10.1016/j.eswa.2010.11.100 28 A. Subasi, Chapter 2-Biomedical Signals, Editor(s): Abdulhamit Subasi, Practical Guide for Biomedical Signals Analysis Using Ma- chine Learning Techniques, Academic Press, 2019, 27–87, doi:10.1016/B978-0-12-817444-9.00002-7 29 J. A. Clancy, D. A. Mary, K. K. Witte, J. P. Greenwood, S. A. Deuchars, J. Deuchars, Non-invasive vagus nerve stimulation in healthy humans reduces sympathetic nerve activity, Brain Stimul., 7 (2014) 6, 871–7, doi:10.1016/j.brs.2014.07.031 30 L. V. Borovikova, S. Ivanova, M. Zhang, H. Yang, G. I. Botchkina, L. R. Watkins, H. Wang, N. Abumrad, J. W. Eaton, K. J. Tracey, Vagus nerve stimulation attenuates the systemic inflammatory re- sponse to endotoxin, Nature, 405 (2000) 6785), 458–62, doi:10.1038/35013070 J. ROZMAN et al.: MEASUREMENT OF BASIC HEMODYNAMICS DURING MULTIPOST TRANSCUTANEOUS ... Materiali in tehnologije / Materials and technology 58 (2024) 3, 309–319 319