AUTOMATIC SCREWING OF CAPS TO SPIKE CONNECTORS ON APD PERITONEAL DIALYSIS LINES Andrej Čebular MIKROIKS d.o.o., Ljubljana, Slovenia Key words: hemodialysis, peritoneal dialysis, APD, CARD, manual caps screwing, automatic caps screwing. Abstract: Around 3 million people worldwide with end stage renal disease regularly need to use one of the established dialysis methods. The most common is haemodialysis but less invasive is peritoneal dialysis where peritoneum takes the role of artificial kidney /1 /. Using APD - Automatic peritoneal Dialysis, PD dialiser executes the whole cycle of successive charging and dialysis fluid discharging during the night. Connection of lines, solutions and patient to PD dialyser is done by use of connectors which are protected with caps that must be unscrewed sterilly before use. In the production of disposable PDL lines the caps are screwed to connectors manually or automatically. Automatic screwing guarantees controlled and repetitive screwing conditions, as well as higher throughput. In the article we describe such a machine that we built. It is fully automatic and needs to be occasionally refilled with material and reset to define new material lot. Process, as well as production, parameters are put-in through user friendly touch screen. Avtomat za vijačenje kapic na konektorje linij za avtomatsko peritonealno dializo Kjučne besede: hemodializa, peritonealna dializa, APD, CAPD, ročno vijačenje kapic, avtomatsko vijačenje kapic, izvleček: Dandanes se je nekaj manj kot tri milijone bolnikov s stalno ledvično odpovedjo prisiljeno redno zatekati k eni od uveljavljenih metod dialize. Gre bodisi za hemodializo, oz.krvno dializo, kjer bolnika trikrat tedensko priključimo na umetno ledvico, ali pa za peritonealno dializo, kjer vlogo (umetnih ) ledvic prevzame bolnikova potrebušnica, peritonej /1/. Pri kontinuirani peritonealni dializi {CAPD - Continuous Ambulatory Peritoneal Dialysis) bolnik štirikrat dnevno ročno prazni in nato polni močno prekrvav-Ijeno trebušno votlino z ustreznimi raztopinami. Pri avtomatski peritonealni dializi ( APD - Automatic peritoneal Dialysis ) pa to funkcijo prevzame PD dializator. Priključitev linije in vrečk z raztopinami na dializator bolnik opravi s pomočjo ustreznih konektorjev. Le-ti so zaščiteni s čepki, ki jih mora bolnik predhodno pazljivo sterilno odviti. V proizvodnji dializnih linij za enkratno upoprabo se vijačenje čepkov na konektorje izvaja ročno ali avtomatsko. Ročno vijačenje ima določene slabosti, kot so: - nekontroliran navor privijanja, ki posledično lahko povzroči težave bolniku pri odvijanju, če so čepki preveč priviti ali pa netesnost linije, če so čepki premalo priviti - dolgotrajno ročno privijanje zaradi ponavljajočih se gibov lahko povzroči bolečine v prstih in rokah delavcev. Avtomatsko privijanje na drugi strani zagotavlja kontrolirane pogoje privijanja in višjo produktivnost. Delovanje avtomata je v veliki meri samodejno, saj gaje potrebno le občasno posluževati z materialom, kakor tudi vnašati proizvodne parametre za posamezne lote materiala. V prispevku opišemo konstrukcijo in delovanje avtomata za privijanje čepkov Po začetnem testiranju je naprava začela delovati v proizvodnji ter dosegla vse planirane parametre : kapaciteto nad 8.000 privijanj/izmeno ter navor privijanja v željenem oknu od 3Ndm do 8Ndm. 1. Introduction Around 3 million people worldwide with end stage renal disease need to regularly use one of the established dialysis methods. The most common is haemodialysis, which needs to be done three to four times a week where the patient is dialysed by use of artificial kidney. Less invasive is peritoneal dialysis where peritoneum takes the role of artificial kidney/1/. Using CAPD - Continuous Ambulatory Peritoneal Dialysis, the patient needs to manually discharge and charge the peritoneal cavity four times a day with suitable solutions. On the other hand with APD - Automatic peritoneal Dialysis, PD dialiser takes over the whole cycle of successive charging and discharging during the night. Connection of lines, solutions and patient to PD dialyser is done by use of suitable connectors, figure 1 and figure 2. These connectors are protected with caps which must be unscrewed sterilely before use. Cap colour defines to which type of dialysis solution the line connector must be connected to. In the production of disposable PDL lines the caps are screwed to connectors manually. Manual screwing has several disadvantages like: uncontrollable screwing torque which may lead to problems during cap unscrewing by patient due to too tight force, or on the other hand, leakage of the whole line due to too light force continuous repetitive manual screwing may cause operator finger and handaches. Cv ■ • n I Fig. 2: a) cap (bottom) and spike (up) separated; b) cap screwed on spil 0,40 - ^ 0,20 - 0,00 - LINE LEAKAGE after 10s at 450mttijHg torque. Ndin Fig. 11: Line leakage versus torque 0 3500 5000 7500 10000 12500 15000 17500 20000 22500 25000 27500 30000 "Integer" value of anlog output Fig. 10: Graphs of torque measurement and voltage output - a) Theoretical and servo controller feedback torque; b) Linearity of the DA converter 3. Results 3.1 Machine operation Machine has been in operation for several months and besides some minor mechanical and software changes it has been performing according to expectations. 3.2 Machine throughput Planned throughput of S.OOOpcs/shift has been reached very soon. Today, the machine throughput is in the range of 9.500pcs/shift. 3.3 Determination of torque window Torque window within which the cap should be screwed on the connector was determined in the following way : a)if minimum torque is applied the line should still not leak through the connector cap By use of graph shown on figure 11 the minimum torque required before the line starts to leak was determined to be 2Ndm. As can be seen on figure 12, line leakage through the connector screwed with less than 2Ndm becomes substantial. A. ' Fig. 12: Cap to screw leakage b)if maximum torque is applied it becomes impossible to unscrew the cap manually without the appropriate tool. Several people were asked to unscrew the caps and subjectively classify the force they had to use. The results are shown in the Table 1. Obviously torque above 8Ndm is already too high and it becomes impossible to unscrew the cap. Table 1, force needed to unscrew the cap, subjective rating torque, Ndm unscrewing, subjective rating 3 easy 4 easy 5 not so easy 6 difficult 7 almost impossible 8 impossible 20 impossible From above experiments the acceptable torque window was defined to be between 3Ndm and 8Ndm with central value around 5Ndm. 3.4 Screwing results Validation of screwing was executed by torque measurement needed to unscrew the cap. To do so we prepared a measurement accessory similar to the actual machine's screwing head used for automatic screwing but with added dynamometer. Histogram of measured torque (automatic CAP screwing) (n=336) K! Relative frequency rNormal distribution (m = 5.2, stdv=0.9) — Torque [Ndm] Fig. 14: Histogram of torque measurements ger and handaches, as well as we obtained good control of screwing process. Long term results of torque measurements show almost 30 % better accuracy of automatic system over the manual assembly. Operator friendly user interface allows easy machine set up and control of main process parameters. Although main goals of machine operation and process control were met, there is still room for further development and research, especially in shortening cycle time and lowering standard deviation of the unscrewing torque. Fig. 13: Torque measurement To execute the measurement the subassembly must be positioned in the holder, the measurement head lowered and the dynamometer arm forced to open the cap. The force is read and the torque needed to unscrew the cap calculated. The data acquired for the torque needed to unscrew the caps and spikes is represented on figure 14. The calculated mean value of the torque is 5,2 Ndm with standard deviation of 0,9 Ndm (based on population of 336 samples). Shape of the histogram is very close to theoretical normal distribution N (5.2;0.9). 4. Conclusion 5. References /1/ I.Šorli, P.Šorli : Kidney and Kidney Desease - Peritoneal Dialysis, Internal Bioiks training manual, 2009 ( Ledvica in bolezni ledvic - peritonealna dializa, interni priročnik za šolanje, Bioiks 2009) /2/ MITSUBISHI ELECTRIC CORPORATION: General-Purpose AC Servo EZMOTION MR-E SuperGeneral-Purpose Interface MODEL MR-E-A-QW003 INSTRUCTION MANUAL, 2008 /3/ Siemens AG: S7-200 Programmable Controller System Manual, 2007 Andrej Čebular MIKROIKS d. o. o., Ljubljana, Siovenia The machine for automatic screwing of caps to spike connectors was constructed and built. Due to its high throughput it successfully replaced several manual operators with which we also avoided problems of long term operator fin-