research article Efficacy of high-energy collimator for sentinel node lymphoscintigraphy of early breast cancer patients Kamran Aryana1, Mohaddeseh Gholizadeh1, Mehdi Momennezhad1, Maryam Naji1, Mohsen Aliakbarian2, Mohammad Naser Forghani3, Ramin Sadeghi1 1 Nuclear Medicine Research Center, Faculty of Medicine, Imam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran 2 Surgical Oncology Research Center, Faculty of Medicine, Imam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran 3 Cancer Research Center, Faculty of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran Radiol Oncol 2012; 46(1): 75-80. Received 22 October 2011 Accepted 8 December 2011 Correspondence to: Assistant Professor Ramin Sadeghi, Nuclear Medicine Research Center, Imam Reza Hospital, Mashhad University of Medical Sciences, Ebn Sina Street. Mashhad, Iran. E-mail: sadeghir@mums.ac.ir Disclosure: No potential conflicts of interest were disclosed. Introduction. Lymphoscintigraphy is an important part of sentinel node mapping in breast cancer patients. Sometimes star shaped artefacts due to septal penetration can be problematic during imaging. In the current study, we evaluated the possibility of high energy (HE) collimators use for lymphoscintigraphy. Patients and methods. Twenty patients with early breast carcinoma were included. Thirty minutes after radiotracer injection (99mTc-antimony sulphide colloid), anterior and lateral images were acquired using a dual head gamma camera equipped with a parallel hole low energy high resolution (LEHR) collimator on one head and HE collimator on another head. All images were reviewed by two nuclear medicine specialists regarding detectability and number of axillary sentinel nodes and presence of star artefact. Results. All images taken by LEHR collimators showed star artefact of the injection site. No image taken by HE collimator showed this effect. In two patients the sentinel node was visible only by HE collimator. Tumour location in both of these patients was in the upper lateral quadrant and both had history of excisional biopsy. In two patients additional sentinel node was visible adjacent to the first one only on the LEHR images. Conclusions. HE collimators can be used for sentinel lymph node mapping and lymphoscintigraphy of the breast cancer patients. This collimator can almost eliminate star-shaped artefacts due to septal penetration which can be advantageous in some cases. However, to separate two adjacent sentinel nodes from each other LEHR collimators perform better. Key words: sentinel node; lymphoscintigraphy; collimator; HEAP; high energy all purpose; low energy high resolution; LEHR Introduction Sentinel node biopsy is the standard method of axillary staging in early breast cancer patients.1 During surgery the sentinel nodes can be detected with two different techniques; alone or in combination: radiotracer and/or blue dye approaches.2 Imaging after radiotracer injection (lymphoscin- itgraphy) was recommended by most guidelines which can help in performing sentinel node biopsy flawlessly.3 Usually after injection of the radiotracer in the breast (in a specific location according to the used protocol), lymphoscintigraphy was performed in different time intervals.4 A major problem while imaging the axilla, was concealment of the sentinel nodes by the activity in the injection FIGURE 1. Low energy high resolution collimator (LEHR) (upper row) as well as high energy (HE) collimator (lower row) images of a patient. Scatterograms are also shown on the right sides of each original image. There is no star shape artefact in the HE collimator images. TABLE 1. The specifications of high energy (HE) and low energy all purpose collimators Low energy all purpose collimator High energy collimator Hole shape Hexagonal Hexagonal Number of holes (*1,000) 148 8 Hole length (mm) 24.05 50.8 Septal thickness (mm) 0.16 2 Hole diameter (mm across the flats) 1.11 3.4 Sensitivity at10 cm (count per minute/pCi) for 99mTc photons 261.5 285.4 Spatial resolution (mm at 10 cm) for 99mTc photons 6.64 12.66 Septal penetration (%) for 99m Tc photons 1.5 Almost none site, scatter photons, as well as star shaped artefacts due to septal penetration.56 To avoid this problem many centres used a lead shield on the injection site with some success.7 Another approach to decrease the above mentioned problem was to use other types of collimators with thicker septa (such as medium energy) instead of usual low energy ones in order to reduce the septal penetration and decrease the star-shaped artifact.5-10 Although high energy (HE) collimators have thicker septa compared to the medium energy collimators, to the extent of our knowledge the efficacy of this kind of collimator for sentinel node mapping has not been evaluated before. In the current study, we evaluated the feasibility and possible advantages of using HE collimator for lymphoscintigraphy of early stage breast cancer patients. Patients and methods 20 patients with the clinical diagnosis of early (stages I and II) breast carcinoma were included in the study. Histological diagnosis of breast cancer was based on the results of core needle biopsy or excisional biopsy. For patients in whom the diagnosis was established by core needle biopsy periareolar intradermal injections of 18.5 MBq (0.5 mCi)/0.2 mL 99mTc-antimony sulphide colloid were used. For patients with history of previous excisional biopsy of the primary lesion two intradermal injections of 18.5 MBq (0.5 mCi)/0.2 mL 99mTc-antimony sulphide colloid at each end of the excisional line were given. Gentle massage was applied to the injection site subsequently for all injections for 1 minute. Anterior and lateral views were acquired 30 minutes after the injection (2 minutes/image, 128x128 matrix, 15% energy window cantered over 140 keV) using a dual head gamma camera (e.cam Siemens), equipped with a parallel hole low energy high resolution (LEHR) collimator on one head and HE energy collimator on another head. The order of imaging was: 1) lateral view with HE, 2) lateral view with low energy, 3) anterior view with HE and 4) anterior view with low energy col-limators. The outline of the patients was acquired simultaneously using the scattered photons as described by Momennezhad et al.11 The specifications of both HE and LEHR collimators are provided in Table 1. Some of the gamma camera specifications are shown in Table 2. Collimator performance was calculated using especial calculator provided by Nuclear Fields Company.12 All images were reviewed by two nuclear medicine specialists regarding: detectability of axillary sentinel nodes, number of visualized sentinel nodes, and presence of star artefact. Semiquantitative evaluation was also performed using ROIs on the injection site and on detected axillary sentinel nodes. Quantitative data (count rates) were expressed as mean ± SD. For comparison of these quantitative data between two sets of images paired sample t tests was used. P values less than 0.05 were considered statistically significant. The study was carried out according to the Declaration of Helsinki. FIGURE 2. Low energy high resolution collimator (LEHR) (upper row) as well as high energy (HE) collimator (lower row) images of a patient. Note that the sentinel node is only visible on the HE collimator images (arrow) due to star artefact on the low energy collimator imaging. The scatterograms are also shown on the left side of each original image. O Lateral Anterior ® Results Table 3 shows the summary of patients' characteristics. All images taken by LEHR collimators showed star artefact of the injection site. No image taken by HE collimator showed this effect (Figure 1). In two Lateral Anterior FIGURE 3. High energy (HE) collimator (A) as well as low energy high resolution col-limator (B) images of the same patient. Note that two separate sentinel nodes are obvious on the low energy high resolution images. These two sentinel nodes are not shown as discrete nodes on the HE collimator images. TABLE 2. Some of the gamma camera specifications used in the current study Specifications Field-of-view (FOV) Diagonal FOV Crystal Size Diagonal Thickness Photomultiplier tubes Total number Type Array Shielding Back Sides Intrinsic spatial resolution FWHM in CFOV FWHM in UFOV Intrinsic energy resolution System spatial resolution without scatter with LEHR collimator at 10 cm 53.3 x 38.7 cm 63.5 cm 59.1 x 44.5 cm 69.2 cm 9.5 cm 59 Bialkali high-efficiency box-type dynodes Hexagonal 9.5 mm 12.7 mm <3.8 mm <3.9 mm <9.9% 7.4 mm FWHM = full width at half maximum; CFOV = central field of view; useful field of view UFOV; LEHR = low energy high resolution collimator TABLE 3. Patients characteristics Age Tumour size (in cm) History of exci-sional biopsy Tumour location Number of detected lymph nodes with LEHR collimator Number of detected lymph nodes with HE collimator 1 31 1.2 No UM 1 1 2 34 2.2 No UL 2 1 3 56 2.3 Yes UL 0 1 4 57 2.4 No UL 0 5 80 3.1 No LL 1 1 6 43 3.6 Yes LM 1 1 7 44 2 Yes UL 1 1 8 36 2.1 Yes Central 2 1 9 47 2.4 No LL 2 1 10 58 1.3 Yes UL 1 1 11 55 1.5 No UL 1 1 12 45 3 Yes LL 1 1 13 45 1.3 Yes UM 1 1 14 78 1.4 No UL 1 1 15 56 2.5 No LL 1 1 16 32 2.2 No LM 1 1 17 36 2.3 No Central 1 1 18 45 1.7 Yes UL 0 1 19 30 1.8 No UL 1 1 20 34 1.9 Yes UL 1 1 TABLE 4. Comparison of mean count rates for injection site as well as sentinel node(s) on low energy high resolution (LEHR) collimator and high energy (HE) collimators ... , ... , Low energy high resolution col- „ . High energy collimator |i^mator ° value Injection site counts on lateral views 9127632±78784 8312982±67565 <0.001 Injection site counts on anterior views 9252343±86772 8811054±76453 <0.001 Sentinel node counts on lateral views 1799±67 1576±55 <0.0001 patients the sentinel node was visible by HE collimator but not LEHR collimator. Tumour location in both of these patients was in the upper lateral quadrant and both had history of excisional biopsy (Figure 2). In two patients, additional sentinel node was visible adjacent to the first one only on the LEHR images (Figure 3). Table 4 compares the counts of injection site as well as sentinel node(s) for HE and LEHR collimators. Discussion Imaging of the axillary sentinel lymph nodes (lym-phoscintigraphy) was recommended by many as a necessary part of breast cancer axillary staging using sentinel node biopsy.1314 This imaging can be somehow challenging especially when the injection site is near the axilla which can obscure the sentinel nodes due to "shine through" and star artifacts.715-17 Sub-areolar injection of the tracer by increasing the distance between injection site and axilla can obviate this problem to some extent. However, a problem still persists especially for tumours located in the upper lateral quadrants particularly in patients with the history of previous excisional biopsy of the primary breast lesion.17 A method to decrease the masking effect of injection site count on the sentinel node(s) was using other types of collimators with less septal penetration.9 For the first time in the literature we used HE collimators for lymphoscintigraphy imaging in the current study. Our results showed better visualization of axillary sentinel lymph nodes in two patients with history of excisional biopsy and the primary tumour location in the upper lateral quadrants by HE collimator. This is most likely due to minimal septal penetration by 140 KeV photons of 99mTc using the HE collimator (minimal septal penetration using HE versus 1.5% using LEHR colli-mators). Injection sites in these two patients were very near the axilla and resulting septal penetration masked the axillary sentinel nodes by LEHR collimator. Our findings were also supported by Tsushima et al. who used medium energy collimators. They reported a decrease in star-shaped artefacts with a better chance of sentinel node visualization when injection site is near the axilla.7 The count rate of the injection site as well as of sentinel nodes was consistently higher using HE collimator as compared to LEHR one in our study. This was most likely due to higher sensitivity of HE collimator for 140 KeV photons compared to LEHR (285.4 counts per minute per |jCi for HE and 261.5 counts per minute per |jCi for LEHR). This effect could also contribute to better sentinel node visualization in the above mentioned patients using HE collimators. Star-shaped artefact which is the result of sep-tal penetration is not the only determinant of sentinel node masking by adjacent high counts. The spatial resolution of the collimators is also of utmost importance in this regard. Several authors reported better performance of high resolution collimators for lymphoscintigraphy.5'61018 We also find the same findings in two of our patients. Due to low resolution of HE collimator, the activity of the adjacent sentinel nodes were merged with each other and they can not be separable as two distinct nodes as they were on the LEHR images. Fenestra et al.10 and Lemstra et al.5 both found the same findings comparing medium energy and low energy collimators and attributed this to a higher spatial resolution. This can be especially true in our study since spatial resolution for 99mTc photons (in mm at 10 cm) is 12.66 for HE and 6.64 for LEHR collima-tors respectively. Conclusions HE collimators can be used for sentinel lymph node mapping and lymphoscintigraphy of the breast cancer patients. This collimator can almost eliminate star-shaped artefacts due to septal penetration which can be advantageous in some cases. However, to separate two adjacent sentinel nodes from each other LEHR collimators perform better. We do not recommend routine use of HE collimators for lymphoscintigraphy due to the above mentioned low resolution. In case of significant star-shaped artefacts which can mask the sentinel nodes in the axillary area on the LEHR images, HE collimators can be helpful. Acknowledgement This study was supported by a grant from vice chancellor of research of Mashhad University of Medical Sciences and is the result of a thesis under the approval number of 88673. 14. Abdollahi A, Jangjoo A, Dabbagh Kakhki VR, Zakavi SR, Memar B, Forghani MN, et al. Factors affecting sentinel lymph node detection failure in breast cancer patients using intradermal injection of the tracer. Rev Esp Med Nucl 2010; 29: 73-7. 15. Tsushima H, Yamanaga T, Shimonishi Y, Ochi H. Usefulness of imaging method without using lead plate for sentinel lymph node scintigraphy. Kaku Igaku 2002; 39: 161-9. 16. Jastrzebski T, Kopacz A, Lass P. Comparison of peritumoral and subareolar injection of Tc99m sulphur colloid and blue-dye for detection of the sentinel lymph node in breast cancer. Nucl Med Rev Cent East Eur 2002; 5: 159-61. 17. Maza S, Thomas A, Winzer KJ, Huttner C, Blohmer JU, Hauschild M, et al. Subareolar injection of technetium-99m nanocolloid yields reliable data on the axillary lymph node tumour status in breast cancer patients with previous manipulations on the primary tumour: a prospective study of 117 patients. Eur J Nucl Med Mol Imaging 2004; 31: 671-5. 18. Mariani G, Moresco L, Viale G, Villa G, Bagnasco M, Canavese G, et al. Radioguided sentinel lymph node biopsy in breast cancer surgery. J Nucl Med 2001; 42: 1198-215. References 1. Eftekhari M, Beiki D, Fallahi B, Arabi M, Memari F, Gholamrezanezhad A, et al. Assessment the diagnostic accuracy of sentinel lymph nodes lymphoscin-tigraphy using Technetium-99m phytate in breast cancer. DARU J Pharm Sci 2009; 17: 83-7. 2. Rucigaj TP, Leskovec NK, Zunter VT. Lymphedema following cancer therapy in Slovenia: a frequently overlooked condition? Radiol Oncol 2010; 44: 244-8. 3. Buscombe J, Paganelli G, Burak ZE, Waddington W, Maublant J, Prats E, et al. Sentinel node in breast cancer procedural guidelines. Eur J Nucl Med Mol Imaging 2007; 34: 2154-9. 4. Sadeghi R, Forghani MN, Memar B, Rajabi Mashhadi MT, Dabbagh Kakhki VR, Abdollahi A, et al. How long the lymphoscintigraphy imaging should be continued for sentinel lymph node mapping? Ann Nucl Med 2009; 23: 50710. 5. Lemstra C, Broersma M, Poot L, Jager PL. Sentinel node detection in patients with breast cancer: low-energy all-purpose collimator or medium-energy collimator? Clin Nucl Med 2004; 29: 609-13. 6. Nunez M, Alonso O, Mut F, Vila R, Beretta M, Illutton BF. Detectability of sentinel nodes in lymphoscintigraphy as a function of collimator type and distance from the injection site: A phantom study. J Nucl Med Technol 2004; 32: 105-6. 7. Tsushima H, Takayama T, Yamanaga T, Kizu H, Shimonishi Y, Kosakai K, et al. Usefulness of medium-energy collimator for sentinel node lymphoscintigra-phy imaging in breast cancer patients. J Nucl Med Technol 2006; 34: 153-9. 8. Tsushima H, Takayama T, Kizu H, Yamanaga T, Shimonishi Y, Kosakai K, et al. Advantages of upright position imaging with medium-energy collimator for sentinel node lymphoscintigraphy in breast cancer patients. Ann Nucl Med 2007; 21: 123-8. 9. Krynyckyi BR, Sata S, Zolty I, Kim DW, Kim SC, Knesaurek K. Sentinel node detection in patients with breast cancer: low-energy all-purpose collimator or medium-energy collimator? Clin Nucl Med 2005; 30: 369-70. 10. Fenestra C, Lansbergen SM, Busemann-Sokole E, Sloof GW. Imaging of the sentinel nodes comparing different collimators. Eur J Nucl Med 2000; 27: 1006. 11. Momennezhad M, Zakavi SR, Dabbagh Kakhki VR, Jangjoo A, Ghavamnasiri MR, Sadeghi R. Scatterogram: a method for outlining the body during lym-phoscintigraphy without using external flood source. Radiol Oncol 2011; 45: 184-8. 12. Design calculator. Nuclear Fields Website: http://www.nuclearfields.com/ collimators-design-calculator.htm. Accessed on 1 June 2011. 13. Somasundaram SK, Chicken DW, Waddington WA, Bomanji J, Ell PJ, Keshtgar MR. Sentinel node imaging in breast cancer using superficial injections: technical details and observations. Eur J Surg Oncol 2009; 35:1250-6.